|title=[[Image:O2k-Protocols.jpg|right|80px|link=O2k-Protocols|O2k-Protocols]] Polarographic oxygen sensors: calibration, accuracy and quality control SOP.
|title=[[Image:O2k-Protocols.jpg|right|80px|link=O2k-Protocols|O2k-Protocols]] Polarographic oxygen sensors: calibration, accuracy and quality control SOP. [[Media:MiPNet06.03 POS-Calibration-SOP.pdf |»Bioblast Access pdf«]]
|info=[[Media:MiPNet06.03 POS-Calibration-SOP.pdf|'''Open Access-MiPNet06.03.pdf''']]; [http://www.bioblast.at/index.php/File:MiPNet06.03_POS-Calibration-SOP.pdf Versions]
|info=[http://www.bioblast.at/index.php/File:MiPNet06.03_POS-Calibration-SOP.pdf Versions]
|authors=OROBOROS
|authors=OROBOROS
|year=2015-05-10
|year=2015-05-10

Revision as of 10:05, 17 May 2015

Publications in the MiPMap
O2k-Protocols
Polarographic oxygen sensors: calibration, accuracy and quality control SOP. »Bioblast Access pdf«

» Versions

OROBOROS (2015-05-10) Mitochondr Physiol Network

Abstract: Gnaiger E (2015) Polarographic oxygen sensors: calibration, accuracy and quality control SOP. Mitochondr Physiol Network 06.03(13):1-8.

High-resolution respirometry critically depends on accurate calibration of the polarographic oxygen sensors which respond to partial oxygen pressure. Expressing the oxygen signal in terms of partial pressure has several advantages, but respiration is expressed in molar units related to biochemical stoichiometries. Conversion of oxygen partial pressure to oxygen concentration requires accurate information on oxygen solubilities in experimental incubation media. Absolute errors up to 10% are commonly found in the established literature. The oxygen solubility of mitochondrial respiration medium MiR06 relative to pure water (oxygen solubility factor, FM) is 0.92, accurately determined (for MiR05, i.e. MiR06 without catalase; MiPNet14.13) at 37 °C and 30 °C. At air saturation at standard barometric pressure (100 kPa) and 37 °C, the partial oxygen pressure is 19.63 kPa, and the oxygen concentration is 190.7 µM in MiR06. A clean and ready-to-use O2k-Chamber is a requirement for saving time and quickly calibrating the polarographic oxygen sensors.

» Product: OroboPOS, Oxygraph-2k, O2k-Catalogue


O2k-Network Lab: AT_Innsbruck_OROBOROS


Labels: MiParea: Instruments;methods 





HRR: Oxygraph-2k, Protocol"Protocol" is not in the list (Oxygraph-2k, TIP2k, O2k-Fluorometer, pH, NO, TPP, Ca, O2k-Spectrophotometer, O2k-Manual, O2k-Protocol, ...) of allowed values for the "Instrument and method" property. 

O2k-SOP 

Further details


cO2* is the oxygen concentration at thermodynamic equilibrium with air under standard conditions of 100 kPa barometric pressure. FM has been determined for MiR05 at 30 °C and 37 °C, and is assumed to be identical at 25 °C.