O2k-OroboPOS-Holder MitoPedia

From Bioblast
Revision as of 15:05, 4 March 2020 by Antunes Diana (talk | contribs) (Created page with "{{MitoPedia |description= |info=OroboPOS-Holder }} {{MitoPedia O2k and high-resolution respirometry |mitopedia O2k and high-resolution respirometry=O2k hardware }} __TOC...")
(diff) ← Older revision | Latest revision (diff) | Newer revision β†’ (diff)


high-resolution terminology - matching measurements at high-resolution


O2k-OroboPOS-Holder MitoPedia

Description

Reference: OroboPOS-Holder


MitoPedia O2k and high-resolution respirometry: O2k hardware 



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



O2k-OroboPOS-Holder: Leaky chamber due to damaged OroboPOS-Holder

  • In order not to experience a leakage after chamber reassembly, make sure that all the parts are in good conditions, especially the OroboPOS-Holder
  • The OroboPOS-Holder may experience some damage due to the not proper alignment with the thread
  • The OroboPOS-Holder may experience some damage due to the not proper alignment with the chamber
  • The OroboPOS-Holder may experience some damage due to frequent and bad performed mounting and dismounting procedures
  • In case your OroboPOS-Holder already presents dents on the surfaces ( see pictures below) you may experience leakage, sanding the edge of the OroboPOS-Holder with sand paper may solve the issue.
OroboPOS-Holder002.jpg
OroboPOS-Holder003.jpg
OroboPOS-Holder004.jpg
  • Left picture: brand new OroboPOS-Holder, no damage on the thred, no damage on the edge
  • Center picture: used OroboPOS-Holder, almost no damage on the thred, the edge is slightly damaged
  • Right picture: heavly used OroboPOS-Holder, the thred is slightly damaged, serious damage on the edge of the aperture








MitoPedia O2k and high-resolution respirometry: O2k hardware