O2k-OroboPOS-Holder: Leaky chamber due to damaged OroboPOS-Holder
- In order not to experience a leakage after chamber reassembly, make sure that all the parts are in good conditions, especially the OroboPOS-Holder
- The OroboPOS-Holder may experience some damage due to the not proper alignment with the thread
- The OroboPOS-Holder may experience some damage due to the not proper alignment with the chamber
- The OroboPOS-Holder may experience some damage due to frequent and bad performed mounting and dismounting procedures
- In case your OroboPOS-Holder already presents dents on the surfaces ( see pictures below) you may experience leakage, sanding the edge of the OroboPOS-Holder with sand paper may solve the issue.
- Left picture: brand new OroboPOS-Holder, no damage on the thred, no damage on the edge
- Center picture: used OroboPOS-Holder, almost no damage on the thred, the edge is slightly damaged
- Right picture: heavly used OroboPOS-Holder, the thred is slightly damaged, serious damage on the edge of the aperture
- For more information, please see: Chamber service Insert O2k Chamber