Difference between revisions of "Oxygen sensor test"
From Bioblast
Line 4: | Line 4: | ||
__TOC__ | __TOC__ | ||
'''The | '''The O<sub>2</sub> sensor test is described as an [[OroboPOS]]''' standard operating procedure: '''Polarographic oxygen sensors: calibration, accuracy and quality control SOP.''' »[[MiPNet06.03 POS-Calibration-SOP]]« | ||
Line 14: | Line 14: | ||
== | == The main criteria for sensor performance == | ||
* [[Raw signal at air saturation]] | * [[Raw signal at air saturation]] | ||
* [[Zero current |Raw signal at zero oxygen]] | * [[Zero current |Raw signal at zero oxygen]] | ||
* [[Signal stability]] | * [[Signal stability]]: drift and noise | ||
* [[Response time of the POS]] | * [[Response time of the POS]] | ||
Revision as of 11:06, 16 May 2015
Oxygen sensor test |
MitoPedia O2k and high-resolution respirometry:
O2k-Open Support
The O2 sensor test is described as an OroboPOS standard operating procedure: Polarographic oxygen sensors: calibration, accuracy and quality control SOP. »MiPNet06.03 POS-Calibration-SOP«
When should an O2-sensor test be performed?
- After switching on the O2k, every day: air calibration and stirrer test.
- Zero oxygen calibration: from time to time over weeks; bracketing zero oxygen calibrations when working at low oxygen.
- After application of a new membrane and O2-sensor service: In some cases, the signal of the OroboPOS improves (higher signal stability, less noise, shorter response time), when the O2k remains switched on over night (O2k-Chambers filled with 70% EtOH).
- During troubleshooting, particularly when switching components between the two chambers, a quick sensor test is performed after each step (stirrer test, sensor signal).
The main criteria for sensor performance
- Raw signal at air saturation
- Raw signal at zero oxygen
- Signal stability: drift and noise
- Response time of the POS
... more about "Oxygen sensor test"
O2k-Open Support +