Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "PH calibration buffers"

From Bioblast
(Created page with "{{MitoPedia |description='''pH calibration buffers''' are prepared to obtain two or more defined pH values for calibration of pH electrodes and pH indicator dyes. }} {{MitoPed...")
ย 
Line 3: Line 3:
}}
}}
{{MitoPedia concepts}}
{{MitoPedia concepts}}
{{MitoPedia methods}}
{{MitoPedia methods
|mitopedia method=Respirometry, Fluorometry
}}
{{MitoPedia O2k and high-resolution respirometry}}
{{MitoPedia O2k and high-resolution respirometry}}
{{MitoPedia topics}}
{{MitoPedia topics
|mitopedia topic=Medium
}}
== Hanks based calibration buffers ==
== Hanks based calibration buffers ==
::: '''Modified hanks buffer'''
::: '''Modified hanks buffer'''

Revision as of 10:12, 8 September 2018


high-resolution terminology - matching measurements at high-resolution


PH calibration buffers

Description

pH calibration buffers are prepared to obtain two or more defined pH values for calibration of pH electrodes and pH indicator dyes.



MitoPedia methods: Respirometry, Fluorometry 



MitoPedia topics: Media for respirometry 

Hanks based calibration buffers

Modified hanks buffer
  • CaCl2 * 2 H2O 0.42 mM, imidazole 2.0 mM, potassium chloride 2.4 mM, magnesium chloride hexahydrate 0.49 mM, sodium chloride 137 mM, pH adjusted to 7.1 at 30ยฐC; see also: [1]

ph adjusted with KOH, HCl to create a buffer series: (pH values at 25 ยฐC): A(3.3), B(6.1), C(6.9 initial measurement showed 6.4), D (6.9), E(7.3), F(7.5), G(10.1). for experiments after MF613 buffer C was further adjusted to 6.6 and G was adjusted to 11.1 ( MF613 showed that buffer G at 37ยฐ was not yet representing the basic endpoint for carboxy-SNARF-1)