POS calibration |
MitoPedia O2k and high-resolution respirometry:
O2k-Open Support
Static POS calibration
Static POS calibration = Two-point calibration of the polarographic oxygen sensor.
DatLab converts the raw signal to oxygen concentrations by applying conversion factors determined during a POS calibration comprising an air calibration and a zero calibration. For each gain a different set of calibration values is necessary. Frequently, assumed hardware problems turn out to be a simple calibration problem. This can easily be detected by observing the Raw signal. Please see the Calibration Manual MiPNet19.01D O2k-Calibration.
Dynamic POS calibration
Dynamic POS calibration = Calibration of the response time of the POS.