Search for defective O2k components |
MitoPedia O2k and high-resolution respirometry:
O2k-Open Support
Description
The 2-chamber design of the O2k helps to search for defective O2k components, by switching components linked to O2k chambers A and B between sides A and B.
- The two chamber design of the O2k has many benefits, including advantages in trouble shooting. By switching components between sides A and B of the O2k, it is usually possible to locate an instrumental problem, finding selectively a specific defective component of the O2k. This component can then be serviced by the user or replaced by Oroboros without the need to ship the entire O2k.
[For more information including how to disassemble the OroboPos https://wiki.oroboros.at/index.php/Oxygen_sensor_test#Troubleshooting]