== NO-sensor calibration ==
== NO-sensor calibration ==


I have a question regarding NO-sensor calibration. What method would you recommend?I just tried with a solution (outside of the Oroborous) recommended from the factory (0.05M H2SO4, 0.02M KI with addition of nitrite in different concentrations) This method worked fine, however, is this possible to do in the oroborous? Is that extremely acidic solution harmful to the POS membranes?Best regards
* I have a question regarding NO-sensor calibration. What method would you recommend? I just tried with a solution (outside of the OROBOROS O2k) recommended from the factory (0.05 M H<sub>2</sub>SO<sub>4</sub>, 0.02 M KI with addition of nitrite in different concentrations). This method worked fine, however, is this possible to do in the O2k? Is that extremely acidic solution harmful to the POS membranes? Best regards ~ [[Schiffer TA |Tomas Shiffer]]
Tomas Shiffer
# Answer: That is the method we always used. It is fine for the O2k and no problem for the PTFE O2 membranes. But I have to admit that the laboratory table got some "scars" :-) from the KI. ~ support@oroboros.at, 10 March 2011
Β 
Answer: Thats the method I always used. Its fine for the oxygraph (and of course very fine for the PTFE O2 membrane) but I have to admit the table I used got some "scars" :-) (from the KI).[[User:Fasching Mario|Fasching Mario]] 11:45, 10 March 2011 (CET)
Β 
Β 
{{#set:Technical service=NO (amperometric) signal}}




== Determining pre-polarisation times ==
== Determining pre-polarisation times ==
Please not that for some sensors WPI states the suggested pre-polarisation time not in the manual of the sensor but only in the manual of the WPI electronics,e.g. in the manual for the [http://www.wpiinc.com/products/physiology/tbr4100/?query=%28name.like.TBR4100.or.misc0.like.TBR4100.or.misc2.like.TBR4100.or.misc3.like.TBR4100%29&xsearch_id=products_search_top&xsearch[0]=TBR4100&back=products| TBR4100] (Appendix A)
* For some sensors WPI states the suggested pre-polarisation time not in the manual of their sensor but only in the manual of the WPI electronics.
Β 
* '''User feedback''' for the WPI HPO-2 sensor: ''The suggested polarization voltage for the WPI H<sub>2</sub>O<sub>2</sub> electrodes is 400 mV, that means -400 mV have to be set in the O2k (see [[MiPNet15.05 NO-Manual]]). I have also contacted with WPI. They said that "The ISO-HPO-2 polarized at 450 mV for 2 hours in 0.1 M PBS solution." But I find 2 hours is not enough. I send this information to you, maybe it is helpful for other customers.
Β 
# '''Comment by OROBOROS''': Based on our experience with the WPI NO electrode and our users experience with the WPI HPO electrode, the required prepolarisation time is longer than the (minimum) cited by WPI. Especially for the first run we suggest to allow for a very long pre-polarisation time (minimum over-night, days for NO) and to observe the slope of the signal. Only when the slope of the signal (not the signal itself !) is constant for a long time (hours), pre-polarisation is finished. This will give an indication for the proper pre-polarisation time in further experiments. It is important to continue the pre-polarisation for a few hours after the slope appears to be constant. In the later phase of pre-polarisation the slope may change very slowly but may still approach closer to zero. This can only be detected by recording the slope for a sufficiently long period of time. ~ support@oroboros.at, 23 January 2014
'''From user feedback''' for the WPI HPO-2 sensor: (The suggested polarization voltage for the the WPI H2O2 electrodes is 400 mV, thats means -400 mV have to be set in the O2k, see [[I MiPNet15.05 NO-Manual]].) I have also contacted with WPI,They said that "The ISO-HPO-2 polarized at 450 mV for 2 hours in 0.1M PBS solution." But I find 2 hours is not enough. I send it to you ,maybe it is helpful for your other customer.
Β 
'''Comment by Oroboros Instruments''': With our experiences from the WPI NO electrode and the users experience with the WPI HPO electrode, it seems advisable to assume that required prepolarisation times will be longer than the (minimum) cited by WPI. Especially for the first run we suggest to allow for a very long pre-polarisation time (minimum over night, days for NO) and to observe the slope of the signal. Only when the slope of the signal (not the signal itself !) is constant for a long time (hours), pre-polarisation is finished. This will give an indication for the proper pre-polarisation time in further experiments. It is important to continue the prepolarisation for a few hours after the slope APPEARS to be constant. In the later phase of pre-polarisation the slope may change very slowly but may still improve (that is get closer to zero). This may easily be missed by not observing the slope for long enough. Β 
[[User:Fasching Mario|Fasching Mario]] 09:23, 23 January 2014 (CET)


{{#set:Technical service=NO (amperometric) signal}}


[[Image:BB-Bioblast.jpg|left|40px|link=Bioblast:About|Bioblast wiki]]
[[Image:BB-Bioblast.jpg|left|40px|link=Bioblast:About|Bioblast wiki]]

Revision as of 07:59, 6 February 2016


                  


O2k-Open Support

Talk:MiPNet15.05 NO-manual



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



NO-sensor calibration

  • I have a question regarding NO-sensor calibration. What method would you recommend? I just tried with a solution (outside of the OROBOROS O2k) recommended from the factory (0.05 M H2SO4, 0.02 M KI with addition of nitrite in different concentrations). This method worked fine, however, is this possible to do in the O2k? Is that extremely acidic solution harmful to the POS membranes? Best regards ~ Tomas Shiffer
  1. Answer: That is the method we always used. It is fine for the O2k and no problem for the PTFE O2 membranes. But I have to admit that the laboratory table got some "scars" :-) from the KI. ~ support@oroboros.at, 10 March 2011


Determining pre-polarisation times

  • For some sensors WPI states the suggested pre-polarisation time not in the manual of their sensor but only in the manual of the WPI electronics.
  • User feedback for the WPI HPO-2 sensor: The suggested polarization voltage for the WPI H2O2 electrodes is 400 mV, that means -400 mV have to be set in the O2k (see MiPNet15.05 NO-Manual). I have also contacted with WPI. They said that "The ISO-HPO-2 polarized at 450 mV for 2 hours in 0.1 M PBS solution." But I find 2 hours is not enough. I send this information to you, maybe it is helpful for other customers.
  1. Comment by OROBOROS: Based on our experience with the WPI NO electrode and our users experience with the WPI HPO electrode, the required prepolarisation time is longer than the (minimum) cited by WPI. Especially for the first run we suggest to allow for a very long pre-polarisation time (minimum over-night, days for NO) and to observe the slope of the signal. Only when the slope of the signal (not the signal itself !) is constant for a long time (hours), pre-polarisation is finished. This will give an indication for the proper pre-polarisation time in further experiments. It is important to continue the pre-polarisation for a few hours after the slope appears to be constant. In the later phase of pre-polarisation the slope may change very slowly but may still approach closer to zero. This can only be detected by recording the slope for a sufficiently long period of time. ~ support@oroboros.at, 23 January 2014