Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "Talk:O2k-Peltier Temperature Control"

From Bioblast
(Created page with "== Very low temperatures == '''Question:''' Is it possible to us the oxygraph with a chamber temperature of 0 °C? Our system is specified down to 4 °C at 25 deg room temperat...")
 
(Replaced content with "Previous Product ID 21020")
 
(16 intermediate revisions by 7 users not shown)
Line 1: Line 1:
== Very low temperatures ==
Previous Product ID 21020
 
'''Question:''' Is it possible to us the oxygraph with a chamber temperature of 0 °C?
 
Our system is specified down to 4 °C at 25 deg room temperature and and down to 2 °C at "lower ambient temperature", see
[http://www.oroboros.at/index.php?id=o2k-specifications| specifications].
Lower ambient temperature does not necessary mean a "cool room". For the 2 deg, an acclimatised room usually will be enough.
While  we can only guarantee the working of the instrument within these  limits, it is possible to go to 0°C and has been done. In general, this  may require a "cool room". To this purpose the main unit itself can be  exposed to temperatures  down to just above 0°C (all the electronic  components are rated at least down to 0°C but obviously 0°C room  temperature has to be avoided due to condensation concerns). Also of  course for very low room temperatures above 0 deg humidity has to be  controlled somehow.
We did a quick test today and could reach 0°C  chamber temperature quite easily at a room temperature of about 14 °C.  To keep this temperature stable the cooling power of the Peltier element  was already working at 95% of maximum, so some kind of cool room will  probable be necessary
 
 
 
{{#set:Technical service=Main unit}}
 
__SHOWFACTBOX__
{{Technical service}}

Latest revision as of 09:00, 6 August 2017

Previous Product ID 21020