Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "Talk:O2k-chamber"

From Bioblast
Line 13: Line 13:
'''Calibration and quality control of the OroboPOS (O2k-SOP)''' in [[MiPNet06.03 POS-Calibration-SOP]].
'''Calibration and quality control of the OroboPOS (O2k-SOP)''' in [[MiPNet06.03 POS-Calibration-SOP]].


Possible instrumental and protocol related causes for fluxes differing between chambers.
Possible instrumental and protocol related causes for fluxes differing between chambers:


* wrong O2 calibration in one chamber
* wrong O2 calibration in one chamber, see [[MiPNet06.03 POS-Calibration-SOP]]
* wrong chamber volume calibration: The total amount of oxygen in the chamber with the smaller volume will be smaller. Therefore the O2 concentration will drop faster resulting in a higher flux.
* wrong chamber volume calibration: The total amount of oxygen in the chamber with the smaller volume will be smaller. Therefore, the O2 concentration will drop faster resulting in a higher flux, see [[MiPNet19.18A O2k-Start]].
* wrong background parameters: wrong background parameters will cause a O2 concentration dependent offset to the flux, the value of the offset will be independent of the absolute value of the flux. E.g. a constant offset of 10 pmol /ml s at ca 100 µM O2 concentration and 20 pmol/ml s at 300 µM O2 concentration could be caused by wrong background parameters, but  a difference of always 20% at very different absolute values of the flux can not be caused by wrong background parameters. In any case it is advisable to check the background parameters ( redo an instrumental O2 background experiment).
* wrong background parameters: wrong background parameters will cause a O2 concentration dependent offset to the flux, the value of the offset will be independent of the absolute value of the flux. E.g. a constant offset of 10 pmol /ml s at ca 100 µM O2 concentration and 20 pmol/ml s at 300 µM O2 concentration could be caused by wrong background parameters, but  a difference of always e.g. 20% at very different absolute values of the flux can not be caused by wrong background parameters. In any case it is advisable to check the background parameters ( redo an instrumental O2 background experiment). See [[MiPNet14.06 InstrumentalBackground]].
* hydrophobic inhibitors
* hydrophobic inhibitors, see [[MiPNet19.03 O2k-cleaning and ISS]]
* biological contamination
* biological contamination, see [[Biological contamination]]


Hardware defects: While hardware defects (sensors, electronics,..) can obviously have many negative effects (noise signal, slow response, no signal,..) it is difficult to see how a real hardware defect can cause a systematic error in flux calculation once it was possible to do a correct calibration of the O2 sensor (at air saturation and at zero oxygen).
Hardware defects: While hardware defects (sensors, electronics,..) can obviously have many negative effects (noise signal, slow response, no signal,..) it is difficult to see how a real hardware defect can cause a systematic error in flux calculation once it was possible to do a correct calibration of the O2 sensor (at air saturation and at zero oxygen).

Revision as of 12:00, 6 November 2014

Previous Product ID 23100-21

Bioblast wiki

Popular Bioblast page

O2k-Chamber has been accessed more than

  • 5,000 times (2014-08-25)

Different fluxes in left and right chmaber

Question: In our experiments we get consistently higher fluxes in chamber A as compared to chamber B.

Answer: To exclude instrumental causes please follow the procedure described in the section Calibration and quality control of the OroboPOS (O2k-SOP) in MiPNet06.03 POS-Calibration-SOP.

Possible instrumental and protocol related causes for fluxes differing between chambers:

  • wrong O2 calibration in one chamber, see MiPNet06.03 POS-Calibration-SOP
  • wrong chamber volume calibration: The total amount of oxygen in the chamber with the smaller volume will be smaller. Therefore, the O2 concentration will drop faster resulting in a higher flux, see MiPNet19.18A O2k-Start.
  • wrong background parameters: wrong background parameters will cause a O2 concentration dependent offset to the flux, the value of the offset will be independent of the absolute value of the flux. E.g. a constant offset of 10 pmol /ml s at ca 100 µM O2 concentration and 20 pmol/ml s at 300 µM O2 concentration could be caused by wrong background parameters, but a difference of always e.g. 20% at very different absolute values of the flux can not be caused by wrong background parameters. In any case it is advisable to check the background parameters ( redo an instrumental O2 background experiment). See MiPNet14.06 InstrumentalBackground.
  • hydrophobic inhibitors, see MiPNet19.03 O2k-cleaning and ISS
  • biological contamination, see Biological contamination

Hardware defects: While hardware defects (sensors, electronics,..) can obviously have many negative effects (noise signal, slow response, no signal,..) it is difficult to see how a real hardware defect can cause a systematic error in flux calculation once it was possible to do a correct calibration of the O2 sensor (at air saturation and at zero oxygen).