Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "Talk:Raw signal of the oxygen sensor"

From Bioblast
Line 13: Line 13:
::::* at Gain 4 : ca. 8 - 10 V (do not use at air saturation, especially if your lab is situated at or near sea level)
::::* at Gain 4 : ca. 8 - 10 V (do not use at air saturation, especially if your lab is situated at or near sea level)
::::* at Gain 2: ca. 4 - 5 V
::::* at Gain 2: ca. 4 - 5 V
::::* at Gain 1: ca. 2 V
::::* at Gain 1: ca. 1 - 3 V


:::: The raw signal observed during calibration at air saturation depends on experimental temperature, barometric pressure (lower values are obtained above sea level), and the sensitivity of the individual POS. The raw signal is converted to oxygen concentration [ยตM or nmol/ml] or partial oxygen pressure [kPa] by calibration of the POS [[MiPNet12.08]].ย  Each gain requires a separate calibration of the oxygen signal. Therefore, after changing the gain you should not only perform a calibration at air saturation (as always before an experiment), but also a new zero oxygen calibration.ย  ย 
:::: The raw signal observed during calibration at air saturation depends on experimental temperature, barometric pressure (lower values are obtained above sea level), and the sensitivity of the individual POS. The raw signal is converted to oxygen concentration [ยตM or nmol/ml] or partial oxygen pressure [kPa] by calibration of the POS [[MiPNet12.08]].ย  Each gain requires a separate calibration of the oxygen signal. Therefore, after changing the gain you should not only perform a calibration at air saturation (as always before an experiment), but also a new zero oxygen calibration.ย  ย 


{{#set:Technical service=O2 sensor|Technical service=O2 signal}}
{{#set:Technical service=O2 sensor|Technical service=O2 signal}}

Revision as of 08:01, 24 April 2018

Bioblast wiki

Popular Bioblast page

MiPNet17.12 Bioblast 2012 has been accessed more than
  • 5,000 times (2016-06-22)



                  


O2k-Open Support

Talk:Raw signal of the oxygen sensor



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



The raw signal of the polarographic oxygen sensor (POS) depends on the gain setting (default is Gain 1 in O2k-Series G, gain 2 in previous Series). The maximum signal is 10 volt. Above 9.99 V, the amplifier is in the saturation range, the apparent signal will remain constant at this value, and the slope will appear to be zero. The raw signal can be modified by choosing an appropriate gain setting.
Typical recorded voltages at air saturation are
  • at Gain 8: offscale ( โ€œ9.99Vโ€)
  • at Gain 4 : ca. 8 - 10 V (do not use at air saturation, especially if your lab is situated at or near sea level)
  • at Gain 2: ca. 4 - 5 V
  • at Gain 1: ca. 1 - 3 V
The raw signal observed during calibration at air saturation depends on experimental temperature, barometric pressure (lower values are obtained above sea level), and the sensitivity of the individual POS. The raw signal is converted to oxygen concentration [ยตM or nmol/ml] or partial oxygen pressure [kPa] by calibration of the POS MiPNet12.08. Each gain requires a separate calibration of the oxygen signal. Therefore, after changing the gain you should not only perform a calibration at air saturation (as always before an experiment), but also a new zero oxygen calibration.