{{Product
{{Technical support}}
|description='''Filter-Cap''': Each [[Fluorescence-Sensor Green]] and [[Fluorescence-Sensor Cyan]] is equipped with a removable Filter-Cap for exchange of optical filters, which is possible independently for optical pathways from the LED and to the photodiode.
|product id=44310-01
|product type=O2k, MultiSensor, O2k-Fluorescence LED2-Module, O2k-Manual
|info=[[O2k-Fluorescence LED2-Module]]
|product image=[[File:Filter-Cap.JPG|180px|right|link=]]
}}
__TOC__
__TOC__
== O2k-Manual ==
{{MitoPedia without banner
|abbr=
|description=[[Image:Filter-Cap.JPG|180px|right]]'''Filter-Cap''': O2k-Fluo LED2-Module (O2k-Series D to G) sensors ([[Fluorescence-Sensor Green]] and [[Fluorescence-Sensor Blue]]) and O2k-FluoRespirometer (O2k-Series H to I) sensors ([[Smart Fluo-Sensor Green]] and [[Smart Fluo-Sensor Blue]]) are equipped with a removable Filter-Cap for exchange of optical filters for the optical pathways from the LED to the sample and from the sample to the photodiode.
|info= [https://www.oroboros.at/index.php/product/filter-cap/ '''Product details and purchase information''']}}
Β 
== Technical Support ==
[[File:Fluorescence-Sensor lettered.jpg|400px|right|link=]]
[[File:Fluorescence-Sensor lettered.jpg|400px|right|link=]]
=== Selecting a Filter Set ===
:::: The Filter-Cap of each sensor can be removed for application of various filter combinations on the same optical sensor.
:::* Selection of a Filter Set: >> [[MiPNet17.05_O2k-Fluorescence_LED2-Module#Fluorophores| Application specific settings]].


=== Mounting a Filter-Cap ===
=== Mounting a Filter-Cap ===
The Filter-Cap of each sensor can be exchanged for application of different filter combinations on the same optical sensor.
:::* '''Applies to''': O2k-Fluo LED2-Module (O2k-Series D to G) Fluorescence Sensor; O2k-FluoRespirometer (O2k-Series H to I) with Smart Fluo-Sensors
* Dismounting: Pull the Filter-Cap straight from the sensor. The Filter-Cap Guide prevents rotational movements. Insert a filter into either or both Filter-Cap windows for the LED or photodiode.
:::* Dismounting: Pull the Filter-Cap straight from the sensor. The Filter-Cap Guide prevents rotational movements. Β 
* Mounting: Align the Filter-Cap with the Filter-Cap Guide (small steel rod) protruding from the sensor. Press the Filter-Cap onto the sensor without rotational movements.
:::* Remove all filters and store them in the filter box labeled for this Filter Set.
:::* Insert the filters form the selected Filter Set: The round filters of each Filter Set fit to the round window of the Filter-Cap and cover the LED, the rectangular filters fit into the rectangular window of the Filter-Cap and cover the photodiode.
:::* Mounting: Hold sensor and filter cap in a vertical position above you. Align the Filter-Cap with the Filter-Cap Guide (small steel rod) protruding from the sensor. Press the Filter-Cap onto the sensor without rotational movements.
Β 
Β 
:::* '''O2k-Guide: Setup - next step [[Fluorescence-Sensor#Connect_Fluorescence-Sensor_to_O2k|Connect Fluorescence-Sensor to O2k]]'''
Β 
==Additional resources==
Β 
::::Β» [[O2k-Fluo Smart-Module]]


::::Β» [[Smart Fluo-Sensor]]


'''Next step in setup: [[Fluorescence-Sensor_Green#Connect_Fluorescence-Sensor_to_O2k|Connect Fluorescence-Sensor to O2k]]'''
{{MitoPedia O2k and high-resolution respirometry
|mitopedia O2k and high-resolution respirometry=O2k hardware, O2k-Open Support, O2k-Respirometry, O2k-FluoRespirometry
}}

Latest revision as of 09:40, 6 May 2021


                  


O2k-Open Support

Filter-Cap



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Description

Filter-Cap.JPG

Filter-Cap: O2k-Fluo LED2-Module (O2k-Series D to G) sensors (Fluorescence-Sensor Green and Fluorescence-Sensor Blue) and O2k-FluoRespirometer (O2k-Series H to I) sensors (Smart Fluo-Sensor Green and Smart Fluo-Sensor Blue) are equipped with a removable Filter-Cap for exchange of optical filters for the optical pathways from the LED to the sample and from the sample to the photodiode.


Reference: Product details and purchase information


Technical Support

Fluorescence-Sensor lettered.jpg

Selecting a Filter Set

The Filter-Cap of each sensor can be removed for application of various filter combinations on the same optical sensor.

Mounting a Filter-Cap

  • Applies to: O2k-Fluo LED2-Module (O2k-Series D to G) Fluorescence Sensor; O2k-FluoRespirometer (O2k-Series H to I) with Smart Fluo-Sensors
  • Dismounting: Pull the Filter-Cap straight from the sensor. The Filter-Cap Guide prevents rotational movements.
  • Remove all filters and store them in the filter box labeled for this Filter Set.
  • Insert the filters form the selected Filter Set: The round filters of each Filter Set fit to the round window of the Filter-Cap and cover the LED, the rectangular filters fit into the rectangular window of the Filter-Cap and cover the photodiode.
  • Mounting: Hold sensor and filter cap in a vertical position above you. Align the Filter-Cap with the Filter-Cap Guide (small steel rod) protruding from the sensor. Press the Filter-Cap onto the sensor without rotational movements.


Additional resources

Β» O2k-Fluo Smart-Module
Β» Smart Fluo-Sensor


MitoPedia O2k and high-resolution respirometry: O2k hardware, O2k-Open Support, O2k-Respirometry, O2k-FluoRespirometry