* 2 Fluorescence-Sensors Green are included in the [[O2k-Fluorescence LED2-Module]].
== O2k-Manual ==
== O2k-Manual ==
[[File:Fluorescence-Sensor lettered.jpg|400px|right|link=]]
[[File:Fluorescence-Sensor lettered.jpg|400px|right|link=]]
* 2 Fluorescence-Sensors Green are included in the [[O2k-Fluorescence LED2-Module]].
=== Mounting a Filter-Cap ===
The Filter-Cap of each sensor can be exchanged for application of different filter combinations on the same optical sensor.
* Dismounting: Pull the Filter-Cap straight from the sensor. The Filter-Cap Guide prevents rotational movements. Insert a filter into either or both Filter-Cap windows for the LED or photodiode.
* Mounting: Align the Filter-Cap with the Filter-Cap Guide (small steel rod) protruding from the sensor. Press the Filter-Cap onto the sensor without rotational movements.


=== Connect ===
=== Connect ===
Connect the cable of the Fluorescence-Sensor to the Fluorescence-Sensor Plug of the Fluorescence-Control Unit. For recording of fluorescence signals, insert the black sensor into the window of the O2k chamber as far as possible, aligning the Sensor-Guide Sector with the Sensor-Guide of the O2k-Front Fixation.
Connect the cable of the Fluorescence-Sensor to the Fluorescence-Sensor Plug of the Fluorescence-Control Unit. For recording of fluorescence signals, insert the black sensor into the window of the O2k chamber as far as possible, aligning the Sensor-Guide Sector with the Sensor-Guide of the O2k-Front Fixation.
* For exchange of filters and mounting the Filter-Cap onto the sensor, see [[Filter-Cap]].

Revision as of 17:35, 11 March 2012

                



Fluorescence-Sensor Green

O2k-Catalogue

Description Fluorescence-Sensor Green: excitation LED 525 nm, photodiode, Filter-Cap attached for Amplex UltraRed measurements.
Product ID 44210-01
Type O2k, MultiSensor, O2k-Fluorescence LED2-Module, O2k-Manual"O2k-Manual" is not in the list (O2k, O2k-Module, O2k-Core, O2k-Fluorometer, O2k-Main Unit, O2k-Assembly Kit, OroboPOS, OroboPOS-Service Kit, ISS, TIP2k, ...) of allowed values for the "Product type" property.
Link O2k-Fluorescence LED2-Module
Image
Fluorescence-Sensor Green.JPG

O2k-Manual

Fluorescence-Sensor lettered.jpg

Connect

Connect the cable of the Fluorescence-Sensor to the Fluorescence-Sensor Plug of the Fluorescence-Control Unit. For recording of fluorescence signals, insert the black sensor into the window of the O2k chamber as far as possible, aligning the Sensor-Guide Sector with the Sensor-Guide of the O2k-Front Fixation.

  • For exchange of filters and mounting the Filter-Cap onto the sensor, see Filter-Cap.