Difference between revisions of "MiPNet14.06 Instrumental O2 background"

From Bioblast
Jump to: navigation, search
Line 8: Line 8:
 
|abstract='''Fasching M, Gnaiger E (2018) O2k Quality Control 2: Instrumental oxygen background correction and accuracy of oxygen flux. Mitochondr Physiol Network 14.6(07):1-14.'''  
 
|abstract='''Fasching M, Gnaiger E (2018) O2k Quality Control 2: Instrumental oxygen background correction and accuracy of oxygen flux. Mitochondr Physiol Network 14.6(07):1-14.'''  
  
'''O2k-Protocols SOP''': Correction for instrumental background oxygen flux is a standard in high-resolution respirometry, automatically performed by DatLab. Background measurements provide a test of instrument function. In the Oroboros O2k, background corrections are usually within a few % of experimental flux over the entire experimental oxygen range. At minimum activities, however, even the small background effects become highly significant and require compliance to standard operating procedures described in this chapter as part of the [[MitoFit Quality Control System]]. This is part 2 of O2k Quality Control.
 
 
:» O2k Quality Control 1 »[[MiPNet06.03 POS-calibration-SOP]]«
 
 
:» Product: [[O2k-FluoRespirometer]], [[Oroboros O2k-Catalogue |O2k-Catalogue]]
 
:» Product: [[O2k-FluoRespirometer]], [[Oroboros O2k-Catalogue |O2k-Catalogue]]
 
:» MitoPedia: [[Instrumental_background_oxygen_flux| Instrumental background oxygen flux]]
 
:» MitoPedia: [[Instrumental_background_oxygen_flux| Instrumental background oxygen flux]]
Line 16: Line 13:
 
|mipnetlab=AT_Innsbruck_Oroboros
 
|mipnetlab=AT_Innsbruck_Oroboros
 
}}
 
}}
 +
{{MitoPedia without banner
 +
|abbr=TIP2k
 +
|description=The '''instrumental O<sub>2</sub> background test''' is part 2 of O2k Quality Control, performed after [[MiPNet06.03 POS-calibration-SOP |air calibration of the oxygen sensor]] when starting an experimental series, and performed at intervals of several weeks, depending on the rogorosity of quality control required in an experimental programme.
 +
|info=
 +
}}
 +
__TOC__
 +
{{Technical support integrated}}
 +
::: '''O2k-Protocols SOP'''
 +
:::: Correction for instrumental background oxygen flux is a standard in high-resolution respirometry, automatically performed by DatLab. Background measurements provide a test of instrument function. In the Oroboros O2k, background corrections are usually within a few % of experimental flux over the entire experimental oxygen range. At minimum activities, however, even the small background effects become highly significant and require compliance to standard operating procedures described in this chapter as part of the [[MitoFit Quality Control System]]. This is part 2 of O2k Quality Control.
 +
 +
:» O2k Quality Control 1 »[[MiPNet06.03 POS-calibration-SOP]]«
 +
 
{{Labeling
 
{{Labeling
 
|area=Respiration, Instruments;methods
 
|area=Respiration, Instruments;methods
Line 21: Line 30:
 
|additional=DatLab, O2k-SOP
 
|additional=DatLab, O2k-SOP
 
}}
 
}}
 
 
{{Technical support integrated}}
 
 
 
[[Category:OroboPedia]]
 

Revision as of 14:06, 7 November 2019

Oroboros Instruments

News         O2k-Products         O2k-Purchase         O2k-Open Support         O2k-Publications         O2k-Workshops and Events         O2k-Network         O2k-Feedback         NextGen-O2k



MiPNet14.06 Instrumental O2 background

Publications in the MiPMap
O2k-Protocols
O2k Quality Control 2: Instrumental oxygen background correction and accuracy of oxygen flux.

» Bioblast pdf » Versions

Oroboros (2018-08-07) Mitochondr Physiol Network

Abstract: Fasching M, Gnaiger E (2018) O2k Quality Control 2: Instrumental oxygen background correction and accuracy of oxygen flux. Mitochondr Physiol Network 14.6(07):1-14.

» Product: O2k-FluoRespirometer, O2k-Catalogue
» MitoPedia: Instrumental background oxygen flux

Keywords: Instrumental O2 background

O2k-Network Lab: AT_Innsbruck_Oroboros


Description

The instrumental O2 background test is part 2 of O2k Quality Control, performed after air calibration of the oxygen sensor when starting an experimental series, and performed at intervals of several weeks, depending on the rogorosity of quality control required in an experimental programme.

Abbreviation: TIP2k

Contents

Template NextGen-O2k.jpg
O2k-Open Support


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 

O2k-Protocols SOP
Correction for instrumental background oxygen flux is a standard in high-resolution respirometry, automatically performed by DatLab. Background measurements provide a test of instrument function. In the Oroboros O2k, background corrections are usually within a few % of experimental flux over the entire experimental oxygen range. At minimum activities, however, even the small background effects become highly significant and require compliance to standard operating procedures described in this chapter as part of the MitoFit Quality Control System. This is part 2 of O2k Quality Control.
» O2k Quality Control 1 »MiPNet06.03 POS-calibration-SOP«


Labels: MiParea: Respiration, Instruments;methods 





HRR: Oxygraph-2k, O2k-Manual 

DatLab, O2k-SOP