{{Technical support}}
|description=In '''O2k configuration''' channels (amperometric and potentiometric) can be switched on/off by selecting the according tick box.
{{MitoPedia without banner
ย 
|description= Configure or modify the settings for the O2k sensors
Power-O2k and numbers for Oxygen sensor, Amp sensor, pX Electrode and pX Reference electrode have to be entered and edited in the mask shown on the left. They are also shown in [[O2k control]] but cannot be edited there.
ย 
ย 


In '''O2k configuration''', channels (amperometric and potentiometric) can be switched on/off by selecting the according tick box. The Power-O2k number (P1, P2, ..) and numbers for O2 sensors, Amp sensors, pX electrodes and pX reference electrodes are entered or edited here. With the [[O2k-FluoRespirometer]] (O2k-Series H and higher), the serial numbers of the [[Smart Fluo-Sensor|Smart Fluo-Sensors]] are shown automatically under [Amperometric, Amp]. The O2k configuration window pops up when DatLab starts and "Connect to O2k" is pressed for the first time. It is also accessible from the menu "Oroboros O2k" and from within the [[O2k control]] and [[Mark statistics - DatLab|Mark statistics]] windows.
}}
}}
{{MitoPedia concepts}}
[[File:Configuration.PNG|right|300px]]
{{MitoPedia methods}}
{{MitoPedia O2k and high-resolution respirometry
{{MitoPedia O2k and high-resolution respirometry
|mitopedia O2k and high-resolution respirometry=DatLab
|mitopedia O2k and high-resolution respirometry=DatLab
}}
}}
{{MitoPedia topics}}
ย 
The O2k configuration window will pop up automatically after connecting with the O2k. If the tick box โ€œSkip configuration at reconnectโ€ is selected, this window will not pop up when reconnecting with the O2k.
== O2k-Series H and higher: Fluorescence sensors serial numbers ==
Communicated by [[Capek O |Ondrej Capek]] 2017-07-11.
ย 
:::: Problem (special case):
::::: (i) O2k-Series H or higher Main-Unit is switched on with fluorescence sensors already connected
::::: (ii) DatLab is started afterwards and "Connect to O2k" clicked - the O2k configuration window is opened
::::: (iii) ''Problem:'' DatLab does not show the corresponding serial numbers of the fluorescence sensors under "Amperometric, Amp"
:::: Solution:
::::: (i) Confirm the O2k configuration window with "OK" or "Cancel" - the [[O2k control]] window is opened
::::: (ii) Open the "O2k configuration" window again (this can be done directly in the [[O2k control]] window by a click on "O2k configuration")
::::: (iii) Now the correct fluorescence sensor serial numbers are shown automatically

Latest revision as of 10:49, 17 November 2022


                  


O2k-Open Support

O2k configuration



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Description

Configure or modify the settings for the O2k sensors

In O2k configuration, channels (amperometric and potentiometric) can be switched on/off by selecting the according tick box. The Power-O2k number (P1, P2, ..) and numbers for O2 sensors, Amp sensors, pX electrodes and pX reference electrodes are entered or edited here. With the O2k-FluoRespirometer (O2k-Series H and higher), the serial numbers of the Smart Fluo-Sensors are shown automatically under [Amperometric, Amp]. The O2k configuration window pops up when DatLab starts and "Connect to O2k" is pressed for the first time. It is also accessible from the menu "Oroboros O2k" and from within the O2k control and Mark statistics windows.

Configuration.PNG

MitoPedia O2k and high-resolution respirometry: DatLab 


O2k-Series H and higher: Fluorescence sensors serial numbers

Communicated by Ondrej Capek 2017-07-11.
Problem (special case):
(i) O2k-Series H or higher Main-Unit is switched on with fluorescence sensors already connected
(ii) DatLab is started afterwards and "Connect to O2k" clicked - the O2k configuration window is opened
(iii) Problem: DatLab does not show the corresponding serial numbers of the fluorescence sensors under "Amperometric, Amp"
Solution:
(i) Confirm the O2k configuration window with "OK" or "Cancel" - the O2k control window is opened
(ii) Open the "O2k configuration" window again (this can be done directly in the O2k control window by a click on "O2k configuration")
(iii) Now the correct fluorescence sensor serial numbers are shown automatically