== O2k-MultiSensor calibration - general ==
[[File:PX calibration 1.PNG|thumb|right|200px]]
[[File:MultiSensor calibration pX.PNG|thumb|right|200px]]
Select a MultiSensor plot and press F5 to open the calibration window, or select the channel from the Calibration menu, or press the pX (or Amp) buttons in the O2k-status line (e.g. left double click on pX calib.).
[[File:PX calibration window.JPG|thumb|right|500px|'''pX calibration''']]
Select a channel and edit Channel label and Unit. In the pX calibration window a drop-down menu allows to select  1, 10^-3, 10^-6 or 10^-9 as Unit, in the Amp calibration window mM, µM or nM can be selected.
Calibrate: Apply changes (recalculates Signal and Slope).
Cancel: Press cancel to exit the calibration window without saving any changes.
== pX channel: two-point calibration - DatLab ==
Linear calibration of pH as a function of recorded voltage is performed by a two-point calibration, using two pH calibration buffers, pX0 and pX1 (for example, pH0 is 7.0 and pH1 is 4.0; consider the actual pH of calibration buffers at experimental temperature).
Example:
1. Add pH calibration buffer 4.0 into respiration chamber, close chamber and insert pH electrode, record data, mark a section of the data when the signal is stable, name Mark pH 4; wash chamber.
2. Add pH calibration buffer 7.0 into respiration chamber, close chamber and insert pH electrode, record data, mark a section of the data when the signal is stable, name Mark pH 7.
3. Select the corresponding two marks in the "Select Mark" column.
4. Press Calibrate.
Note: for the calculation of proton fluxes you need to convert pH changes into proton concentrations. At present this needs to be done in a spreadsheet programm, i.e. outside of DatLab.
<br clear=all>
{{MitoPedia O2k and high-resolution respirometry
{{MitoPedia O2k and high-resolution respirometry
|mitopedia O2k and high-resolution respirometry=DatLab
|mitopedia O2k and high-resolution respirometry=DatLab
}}
}}

Latest revision as of 11:14, 25 March 2022


                  


O2k-Open Support

PX calibration - DatLab



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Description

pX calibration


Reference: MiPNet19.18D_O2k-Series_G_and_DatLab_6:_Calibration, MiPNet08.16 pH calibration

Under construction 
Calibration-menu.png

O2k-MultiSensor calibration - general

PX calibration 1.PNG
MultiSensor calibration pX.PNG

Select a MultiSensor plot and press F5 to open the calibration window, or select the channel from the Calibration menu, or press the pX (or Amp) buttons in the O2k-status line (e.g. left double click on pX calib.).

pX calibration

Select a channel and edit Channel label and Unit. In the pX calibration window a drop-down menu allows to select 1, 10^-3, 10^-6 or 10^-9 as Unit, in the Amp calibration window mM, µM or nM can be selected.

Calibrate: Apply changes (recalculates Signal and Slope).

Cancel: Press cancel to exit the calibration window without saving any changes.

pX channel: two-point calibration - DatLab

Linear calibration of pH as a function of recorded voltage is performed by a two-point calibration, using two pH calibration buffers, pX0 and pX1 (for example, pH0 is 7.0 and pH1 is 4.0; consider the actual pH of calibration buffers at experimental temperature).

Example:

1. Add pH calibration buffer 4.0 into respiration chamber, close chamber and insert pH electrode, record data, mark a section of the data when the signal is stable, name Mark pH 4; wash chamber.

2. Add pH calibration buffer 7.0 into respiration chamber, close chamber and insert pH electrode, record data, mark a section of the data when the signal is stable, name Mark pH 7.

3. Select the corresponding two marks in the "Select Mark" column.

4. Press Calibrate.

Note: for the calculation of proton fluxes you need to convert pH changes into proton concentrations. At present this needs to be done in a spreadsheet programm, i.e. outside of DatLab.


MitoPedia O2k and high-resolution respirometry: DatLab