- Lower O2 limit - DatLab + (A '''Lower O2 limit [ยตM]''' can be defined โฆ A '''Lower O2 limit [ยตM]''' can be defined for each O2k-chamber, to trigger an automatic warning when the experimental O<sub>2</sub> concentration drops below this limit. It reminds the user that re-oxygenation of the O2k-chamber may be required. For the lower O<sub>2</sub> concentration limit, the [[critical oxygen pressure |critical oxygen concentration]] should be considered, which differs between isolated mitochondria, large cells, and permeabilized muscle fibers. A higher limit should be chosen when high oxygen flux is expected, e.g. prior to uncoupler titration. A lower limit is acceptable prior to inhibition of respiration causing low oxygen flux.ptable prior to inhibition of respiration causing low oxygen flux.)