Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "Talk:Oroboros Ecosystem Agenda 2020"

From Bioblast
Line 8: Line 8:
:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Red-loss of data.png|40px|Warning: loss of data]] '''Is the noise of the ‘O2 slope neg’ acceptable?'''
:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Red-loss of data.png|40px|Warning: loss of data]] '''Is the noise of the ‘O2 slope neg’ acceptable?'''
::::::: If you are not certain that the noise of the ‘O2 slope neg’ is within a suitable range or you want to confirm that your experimental data will not be affected, see ''further details'': [[MiPNet06.03 POS-calibration-SOP]]
::::::: If you are not certain that the noise of the ‘O2 slope neg’ is within a suitable range or you want to confirm that your experimental data will not be affected, see ''further details'': [[MiPNet06.03 POS-calibration-SOP]]
:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Yellow-take care.png|40px|Take care]] '''Did you know that you can use the [[TIP2k-Module|Titration-Injection microPump (TIP2k)]] to perform an instrumental O<sub>2</sub> background on the [[O2k-sV-Module|Small O2k-Chamber]] (0.5 mL)?'''
::::::: By adjusting the dithionite concentration to  2.5 mM, instrumental O2 background can be performed with the standard settings on the [[O2k-sV-Module|Small O2k-Chamber]] (0.5 mL).  ''Further details in'' [[Dithionite]] and »[[MiPNet14.06 Instrumental O2 background]]

Revision as of 12:52, 4 March 2020

For discussion

Here we discuss suggestions for the Oroboros Ecosystem Agenda 2020
O2k-Open Support Take care Do you perform a stirrer test during oxygen calibration?
In DatLab 7.4, instrumental DL-Protocols guide you automatically towards performing the stirrer test in the process of air calibration. It shows immediately, if your OroboPOS has a satisfactory response time. Further details in MitoPedia: »Oxygen calibration - DatLab
O2k-Open Support Warning: loss of data Is the noise of the ‘O2 slope neg’ acceptable?
If you are not certain that the noise of the ‘O2 slope neg’ is within a suitable range or you want to confirm that your experimental data will not be affected, see further details: MiPNet06.03 POS-calibration-SOP
O2k-Open Support Take care Did you know that you can use the Titration-Injection microPump (TIP2k) to perform an instrumental O2 background on the Small O2k-Chamber (0.5 mL)?
By adjusting the dithionite concentration to 2.5 mM, instrumental O2 background can be performed with the standard settings on the Small O2k-Chamber (0.5 mL). Further details in Dithionite and »MiPNet14.06 Instrumental O2 background