== Aug 2023 ==
*'''Aug- 14 Mo'''
:{|
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Blue-info.png|25px|Info: improve]]
| '''What do you do if values, symbols, and/or units are not correctly displayed in the DatLab software?'''<br>
It is possible to retrieve the values, symbols and/or correct units by updating the Windows control panel settings. <br>
- Β» '''[[DatLab_error_messages#DatLab_error:_problems_with_text_in_DatLab| Troubleshooting for problems with text in DatLab]]'''Β«<br>
- communicated by Β»[[Cecatto Cristiane]]Β« and Β»[[Timon-Gomez Alba]]Β«
|}
== Jul 2023 ==
== Jul 2023 ==
*'''Jul- 17 Mo'''
*'''Jul- 17 Mo'''

Revision as of 08:53, 14 August 2023

Currently Answered Questions from the Β»Oroboros Ecosystem AgendaΒ«

Aug 2023

  • Aug- 14 Mo
O2k-Open SupportInfo: improve What do you do if values, symbols, and/or units are not correctly displayed in the DatLab software?

It is possible to retrieve the values, symbols and/or correct units by updating the Windows control panel settings.
- Β» Troubleshooting for problems with text in DatLabΒ«
- communicated by Β»Cecatto CristianeΒ« and Β»Timon-Gomez AlbaΒ«

Jul 2023

  • Jul- 17 Mo
O2k-Open SupportInfo: take care Are you planning to perform several measurements in one day using the same O2k-chamber?

To avoid any possible contamination by inhibitors in the subsequent measurements, the O2k-chamber needs to be cleaned each time with the standard protocol AfterUse or AfterUse_inhibitors.
- Β»O2k-cleaning - Sections 2.3 and 2.4Β«
- communicated by Β»Timon-Gomez AlbaΒ« and Β»Cecatto CristianeΒ«

May 2023

  • May- 09 Tu
O2k-Open SupportInfo: take care Does your polarographic oxygen sensor (POS) need a deep clean?

A POS with a very dark anode might need several rounds of ammonia cleaning, possibly including prolonged exposure overnight to restore performance specifications.
- Β»POS serviceΒ«
- communicated by Β»Cecatto CristianeΒ« and Β»Leo ElettraΒ«

Mar 2023

  • Mar- 28 Tu
O2k-Open SupportInfo: improve Unusual stirrer bar rotation?

A β€œjumping” stirrer or a slow rotation of the stirrer bar may indicate a problem either with the stirrer itself or with the cleaning of the glass chamber.
- Β»Stirrer bar - technical supportΒ«
- communicated by Β»Baglivo EleonoraΒ« and Β»Grings MateusΒ«


  • Mar- 21 Tu
O2k-Open SupportWarning: loss of data Did you know that a humid environment and/or liquid spillage may lead to a high raw POS signal [V]?

See how to clean and dry the electrical connections of the OroboPOS-Connector.
- Β»Cleaning the electrical connectionsΒ«
- communicated by Β»Cecatto CristianeΒ« and Β»Baglivo EleonoraΒ«

Feb 2023

  • Feb- 23 Th
O2k-Open SupportInfo: improve Add catalase when measuring Complex IV (CIV) activity in permeabilized fibers (pfi)

Pfi have a high content of iron-containing proteins, which accelerate the autoxidation of tetramethyl-p-phenylenediamine (TMPD) in the CIV assay. Catalase prevents the accumulation of H2O2 generated by this autoxidation.
- Β»Chemical BackgroundΒ« and Β»Complex IVΒ«
- communicated by Β»Cecatto CristianeΒ« and Β»Schmitt SabineΒ«


  • Feb- 16 Th
O2k-Open SupportInfo: improve Do you observe simultaneous fluctuations in the O2 slope neg. and the Peltier power?

This can indicate fluctuations in the room temperature or electrical supply. Controlling the temperature or changing the electrical supply may resolve this problem.
- Β»TroubleshootingΒ«
- communicated by Β»Cecatto CristianeΒ« and Β»Schmitt SabineΒ«