::: '''Volume calibration of the [[O2k-Chamber]]'''
::: '''Volume calibration of the [[O2k-Chamber]]'''
::::* Loosen the stopper fixation rings (A) and (B) with the L-shaped Allen key (OroboPOS-Service Kit).
::::* Loosen the stopper fixation rings (A) and (B) with the Allen key.
::::* Follow the instructions in the O2k-Manual: [[MiPNet22.11 O2k-FluoRespirometer manual]]
::::* Follow the instructions in the O2k-Manual: [[MiPNet22.11 O2k-FluoRespirometer manual]]
::::* Tighten the screws of the fixation rings (A) and (B) gently with the Allen key to ensure that the calibration ring cannot be displaced. Tightening the screw with too much force may damage the thread of the calibration ring. Check for a sufficient fixation by trying to rotate the calibration ring.
::::* Tighten the screws of the fixation rings (A) and (B) gently with the Allen key to ensure that the calibration ring cannot be displaced. Tightening the screw with too much force may damage the thread of the calibration ring. Check for a sufficient fixation by trying to rotate the calibration ring.


[[Image:Volume calibration ring A+B aluminum for Stopper.jpg|right|400px]]
[[Image:Volume calibration ring A+B aluminum for Stopper.jpg|right|400px]]
== Innovation ==
:::: Replacing the (white) PVDF Volume-Calibration Ring with the (black) aluminum Volume-Calibration Ring allows for more force applied to tighten the ring, thus reducing the risk of a loose fit and the associated danger of altering the calibrated position.
-->
[[Category:Technical service]]

Revision as of 11:19, 6 November 2019

                



Volume-Calibration Ring sV

O2k-Catalogue

Description Volume-Calibration Ring sV (A or B), white PVDF, attached to PEEK O2k-Stopper sV.
Product ID 3432A-01 or 3432B-01
Type O2k, Stopper
Link
Image

Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Technical support

Volume calibration of the O2k-Chamber
  • Loosen the stopper fixation rings (A) and (B) with the Allen key.
  • Follow the instructions in the O2k-Manual: MiPNet22.11 O2k-FluoRespirometer manual
  • Tighten the screws of the fixation rings (A) and (B) gently with the Allen key to ensure that the calibration ring cannot be displaced. Tightening the screw with too much force may damage the thread of the calibration ring. Check for a sufficient fixation by trying to rotate the calibration ring.
Volume calibration ring A+B aluminum for Stopper.jpg