:::»'''O2k manual titrations'''  [[MiPNet09.12 O2k-Titrations]]
:::»'''O2k manual titrations'''  [[MiPNet09.12 O2k-Titrations]]
::::* Titration volume ('''2-mL O2k-chamber'''): 5 µL using a 25 µL Hamilton syringe.
::::* Titration volume ('''0.5-mL O2k-chamber'''): 1.25 µL using a 10 µL Hamilton syringe.
::::* Final concentration: 5.5 mM (1 g/L)
::::* Titration volume ('''2-mL O2k-chamber'''): 10 µL using a 25 µL Hamilton syringe.
::::* Titration volume ('''0.5-mL O2k-chamber'''): 2.5 µL using a 10 µL Hamilton syringe.
::::* Final concentration: 11 mM (2 g/L)
<!--


::::* Titration volume: 5 µL using a 25 µL syringe
::::* Titration volume: 5 µL using a 25 µL syringe
Line 35: Line 44:
::::* Titration volume: 10 µL using a 25 µL syringe
::::* Titration volume: 10 µL using a 25 µL syringe
::::* Final conc. in 2 mL O2k-chamber: 11 mM (2 g/L)
::::* Final conc. in 2 mL O2k-chamber: 11 mM (2 g/L)
-->
{{MitoPedia concepts}}
{{MitoPedia concepts}}
{{MitoPedia methods}}
{{MitoPedia methods}}

Revision as of 10:29, 22 October 2020


high-resolution terminology - matching measurements at high-resolution


Glucose

Description

Glucose, also known as D-glucose or dextrose, is a monosaccharide and an important carbohydrate in biology. Cells use it as the primary source of energy and a metabolic intermediate.

Abbreviation: Glc

Reference: MiPNet15.09. Yeast: HRR Reference Assay


Application in HRR

Glc: Glucose (D-(+)-Glucose, C6H1206); Sigma G 8270, 100 g, store at RT; FW = 180.16


Preparation of 2 M stock solution (dissolved in H2O):
  1. weigh 360.32 mg of Glucose
  2. dissolve 1 mL H2O in 1.5 mL Eppendorf
»O2k manual titrations MiPNet09.12 O2k-Titrations
  • Titration volume (2-mL O2k-chamber): 20 µL using a 50 µL Hamilton syringe.
  • Titration volume (0.5-mL O2k-chamber): 5 µL using a 25 µL Hamilton syringe.
  • Final concentration: 20 mM


Preparation of 2.2 M stock solution (dissolved in H2O):
  1. weigh 79.2 mg of Glucose
  2. dissolve 0.2 mL H2O in 0.5 mL Eppendorf
»O2k manual titrations MiPNet09.12 O2k-Titrations
  • Titration volume (2-mL O2k-chamber): 5 µL using a 25 µL Hamilton syringe.
  • Titration volume (0.5-mL O2k-chamber): 1.25 µL using a 10 µL Hamilton syringe.
  • Final concentration: 5.5 mM (1 g/L)
  • Titration volume (2-mL O2k-chamber): 10 µL using a 25 µL Hamilton syringe.
  • Titration volume (0.5-mL O2k-chamber): 2.5 µL using a 10 µL Hamilton syringe.
  • Final concentration: 11 mM (2 g/L)






MitoPedia topics: Substrate and metabolite