{{MitoPedia
{{Technical support}}
{{MitoPedia without banner
|abbr=F10
|abbr=F10
|description=The chambers of the [[OROBOROS O2k|Oroboros O2k]] are illuminated by an internal LED. The '''illumination''' is switched on and off in [[DatLab]] during the experiment by pressing [F10]. This illumination must be distinguished from light introduced into the chambers by LEDs for the purpose of spectrophotometric and fluorometric measurements. For these, the internal illumination must be switched off.
|description=The chambers of the [[OROBOROS O2k|Oroboros O2k]] are illuminated by an internal LED. The '''illumination''' is switched on and off in [[DatLab]] during the experiment by pressing [F10]. This illumination must be distinguished from light introduced into the chambers by LEDs for the purpose of spectrophotometric and fluorometric measurements. For these, the internal illumination must be switched off.
|info=[[MiPNet22.11 O2k-FluoRespirometer manual]]
|info=[[MiPNet22.11 O2k-FluoRespirometer manual]]
}}
}}
{{MitoPedia O2k and high-resolution respirometry
|mitopedia O2k and high-resolution respirometry=DatLab
}}
__TOC__
__TOC__
{{Template:Technical support integrated}}
{{Template:Technical support integrated}}
Line 15: Line 12:
::::* The term '''Light''' (on/off) was used for '''Illumination''' (on/off).
::::* The term '''Light''' (on/off) was used for '''Illumination''' (on/off).


== Troubleshooting ==
== The illumination in both O2k-Chambers does not switch ON ==


::::'''Problem:''' Our O2k chamber lights went off. I couldn’t find the light bulbs in the accessories page on the website. Is there a way to buy the light bulbs and change ourselves?
::::Ana Lira
::::Provided by Ana Lira,  ana-lira@uiowa.edu
::::'''Question:''' Our O2k chamber lights went off. I couldn’t find the light bulbs in the accessories page on the website. Is there a way to buy the light bulbs and change ourselves? (2019-09-16)
:::: '''Answer:'''
:::: '''Answer:'''
:::::* First, check that you did not accidentally press F10 or switch off the illumination in the O2k control window [F7].
:::::* First, please confirm that you did not accidentally press F10 or switch off the illumination in the O2k control window [F7].
:::::* If not, it is suggestes to switch off and on the illumination  using F10 or in the ´O2k control´ window and switch on and off the instrument.
:::::* If not, please perform the following quality control tests:
:::::# Switch off and on the illumination  using F10 or in the ‘O2k control’ window [F7];
:::::# Switch off and on the instrument and check if the illumination is restored.
:::::* If these procedures do not solve the problem, the O2k should be shipped for electronic repair. Please contact our [https://wiki.oroboros.at/index.php/O2k-Open_Support| customer support] to analyse your problem and provide instructions before shipment.
 
{{Template:Oroboros QM}} [[Gnaiger Erich]] 2020-09-07


:::::* If these procedures does not solve the problem, the O2k should be shipped to our electronic workshop for repairment. Please contact our [https://wiki.oroboros.at/index.php/O2k-Open_Support| customer support] to analyse your problem before shippment.
{{MitoPedia O2k and high-resolution respirometry
|mitopedia O2k and high-resolution respirometry=DatLab
}}

Latest revision as of 11:51, 25 March 2022


                  


O2k-Open Support

Illumination



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Description

The chambers of the Oroboros O2k are illuminated by an internal LED. The illumination is switched on and off in DatLab during the experiment by pressing [F10]. This illumination must be distinguished from light introduced into the chambers by LEDs for the purpose of spectrophotometric and fluorometric measurements. For these, the internal illumination must be switched off.

Abbreviation: F10

Reference: MiPNet22.11 O2k-FluoRespirometer manual


Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Previouls DatLab versions

Up to DatLab 6:
  • The term Light (on/off) was used for Illumination (on/off).

The illumination in both O2k-Chambers does not switch ON

Ana Lira
Question: Our O2k chamber lights went off. I couldn’t find the light bulbs in the accessories page on the website. Is there a way to buy the light bulbs and change ourselves? (2019-09-16)
Answer:
  • First, please confirm that you did not accidentally press F10 or switch off the illumination in the O2k control window [F7].
  • If not, please perform the following quality control tests:
  1. Switch off and on the illumination using F10 or in the ‘O2k control’ window [F7];
  2. Switch off and on the instrument and check if the illumination is restored.
  • If these procedures do not solve the problem, the O2k should be shipped for electronic repair. Please contact our customer support to analyse your problem and provide instructions before shipment.
Quality management

Oroboros quality management from the O2k to NextGen-O2k
Gnaiger Erich 2020-09-07


MitoPedia O2k and high-resolution respirometry: DatLab