Difference between revisions of "OroboPOS-Connector"

From Bioblast
Line 22: Line 22:
::::# Unscrew the OroboPOS and check the electrical connections (the thread of the OroboPOS and the gold pin of the OroboPOS-Connector).
::::# Unscrew the OroboPOS and check the electrical connections (the thread of the OroboPOS and the gold pin of the OroboPOS-Connector).
::::# Remove any contaminating grease, salt crystals and moisture with a dry, fine paper cloth.
::::# Remove any contaminating grease, salt crystals and moisture with a dry, fine paper cloth.
::::# If the thread of the OroboPOS and the gold pin of the OroboPOS-Connector are discolored or appear stained, clean repeatedly with a fine paper cloth moistened with 99.9 % ethanol until no more traces appear on the paper cloth used for cleaning.
::::# If the thread of the OroboPOS and the gold pin of the OroboPOS-Connector are discolored or appear stained, clean repeatedly with a fine paper cloth moistened with 99.9 % ethanol until no more traces appear on the paper cloth used for cleaning. It is important to only moisten the paper cloth to avoid EtOH getting into joints which might cause a high leak current. Liquids in joints are difficult to remove and, in many cases, require service by the manufacturer.
::::* Always let the connections (OroboPOS head and OroboPOS-Connector) dry well before use. If the O2k is used in a humid environment or a paper cloth moistened in water was used for cleaning the connection, the POS-Connector and OroboPOS can be dried in an oven at up to 60 ยฐC for 1-2 h. Follow the procedure with an [[Oxygen sensor test]] as a quality control of proper OroboPOS-Connector and OroboPOS performance.
::::* Always let the connections (OroboPOS head and OroboPOS-Connector) dry well before use. If the O2k is used in a humid environment or a moistened paper cloth was used for cleaning the connections, the POS-Connector and OroboPOS should be dried in an oven at up to 60 ยฐC for 1-2 h. Follow the procedure with an [[Oxygen sensor test]] as a quality control of proper OroboPOS-Connector and OroboPOS performance.


== Replacing the OroboPOS-Connector ==
== Replacing the OroboPOS-Connector ==

Revision as of 16:47, 7 November 2023


                  


O2k-Open Support

OroboPOS-Connector



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Description

POS connector and cable connection.jpg

OroboPOS-Connector (blue POM), with male connection to OroboPOS sensor head (POS) and with cable and male plug fitting into O2k-Main Unit.


Reference: Product details and purchase information

Functions

The OroboPOS-Connector has several different functions:
  • It mechanically presses the POS against the glass chamber, ensuring a tight fit.
  • It connects the POS electronically to the Main unit transmitting the signal.
  • In O2k Series A-C it also performs the current to voltage conversion / amplification. For Series D and upwards this function has been incorporated into the Main unit. For this reason OroboPOS connectors for Series A-C and for Series D are NOT interchangeable. The OroboPOS-Connector from O2k-series A-C must only be plugged into or unplugged from the O2k main unit with the O2k switched off.
The OroboPOS-Connector is sensitive to damage by ESD when disconnected from the O2k-Main Unit and when the POS head is removed. Further details: MiPNet14.01_ESD-damage.

Cleaning the body of the OroboPOS-Connector

When experimental media or electrolyte (during service of the polarographic oxygen sensor OroboPOS) leak into the mechanical part of the OroboPOS-Connector, operation of the spring mechanism may be restricted. In such cases, the body of the OroboPOS-Connector should be washed with distilled water. To do so, screw the OroboPOS head onto the OroboPOS-Connector. Make sure that the plug connecting to the O2k-Main Unit does not get wet (protect with parafilm). Rinse the body of the OroboPOS-Connector with distilled water, pulling the spring-loaded part at the base several times. Salt crystals and media clogging the mechanical part are thus washed out. Wipe off the water while again pulling down the spring-loaded part.

Cleaning the electrical connections

  1. Unscrew the OroboPOS and check the electrical connections (the thread of the OroboPOS and the gold pin of the OroboPOS-Connector).
  2. Remove any contaminating grease, salt crystals and moisture with a dry, fine paper cloth.
  3. If the thread of the OroboPOS and the gold pin of the OroboPOS-Connector are discolored or appear stained, clean repeatedly with a fine paper cloth moistened with 99.9 % ethanol until no more traces appear on the paper cloth used for cleaning. It is important to only moisten the paper cloth to avoid EtOH getting into joints which might cause a high leak current. Liquids in joints are difficult to remove and, in many cases, require service by the manufacturer.
  • Always let the connections (OroboPOS head and OroboPOS-Connector) dry well before use. If the O2k is used in a humid environment or a moistened paper cloth was used for cleaning the connections, the POS-Connector and OroboPOS should be dried in an oven at up to 60 ยฐC for 1-2 h. Follow the procedure with an Oxygen sensor test as a quality control of proper OroboPOS-Connector and OroboPOS performance.

Replacing the OroboPOS-Connector

If you need to replace the Orobo-POS Connector, an OroboPOS-Connector Service may be possible.

Additional resources

ยป O2k-Manual