|description='Zero calibration is together with [[air calibration]] one of the twoΒ  steps of the [[POS calibration]].Β  It is done at [[closed chamber]] after all the oxygen has been removedΒ  by the addition of [[dithionite]], see [[MiPNet06.03 POS-Calibration-SOP]]. Unlike airΒ  calibration it is not necessary to perform a zero calibration each day.
|description='Zero calibration is together with [[air calibration]] one of the twoΒ  steps of the [[POS calibration | OroboPOS calibration]].Β  It is performed in theΒ  [[closed chamber]] after all the oxygen has been removedΒ  by the addition of [[dithionite]], see [[MiPNet06.03 POS-Calibration-SOP]]. Unlike airΒ  calibration it is not necessary to perform a zero calibration each day.
|info=[[D MiPNet12.08 O2k-Calibration]]
|info=[[D MiPNet12.08 O2k-Calibration]]
}}
}}
Line 7: Line 7:
|mitopedia method=Respirometry
|mitopedia method=Respirometry
}}
}}
Β 
{{MitoPedia topics}}
== Zero calibration in the O2k ==
== Zero calibration in the O2k ==



Revision as of 22:14, 17 April 2014


high-resolution terminology - matching measurements at high-resolution


Zero calibration

Description

'Zero calibration is together with air calibration one of the two steps of the OroboPOS calibration. It is performed in the closed chamber after all the oxygen has been removed by the addition of dithionite, see MiPNet06.03 POS-Calibration-SOP. Unlike air calibration it is not necessary to perform a zero calibration each day.

Abbreviation: R0

Reference: D MiPNet12.08 O2k-Calibration


MitoPedia methods: Respirometry 



Zero calibration in the O2k

A zero calibration should be performed

A zero calibration has to be performed

  • when a new or serviced POS is used for the first time
  • when a new membrane was applied to the sensor
  • when the gain is changed
  • when a new POS connector is used



Template:Technical service

... more about "Zero calibration"
'Zero calibration is together with air calibration … 'Zero calibration is together with air calibration one of the two steps of the OroboPOS calibration. It is performed in the closed chamber after all the oxygen has been removed by the addition of dithionite, see MiPNet06.03 POS-Calibration-SOP. Unlike air calibration it is not necessary to perform a zero calibration each day.ry to perform a zero calibration each day. +