|abbr=I<sub>O2</sub>
{{MitoPedia without banner
|description=Oxygen flow.
|abbr=''I''<sub>O2</sub> [mol·s<sup>-1</sup>] or [mol·s<sup>-1</sup>·x<sup>-1</sup>]
|type=General
|description=Respiratory '''oxygen flow''' is the oxygen consumption per total [[system]], which is an [[extensive quantity]]. [[Flow]] is advancement of a transformation in a system per time [mol·s<sup>-1</sup>], when 'system' is defined as the experimental system (e.g. an open or closed chamber). Flow is distinguished from the size-specific quantity [[flux]] obtained by normalization of flow per volume of the experimental system [mol·s<sup>-1</sup>·m<sup>-3</sup>]. An experimental object, e.g. a living cell, may be considered as the 'experimental system'. Then oxygen flow per cell has the unit [mol·s<sup>-1</sup>·x<sup>-1</sup>], where [x] is the [[elementary unit]] for a [[count]]. Oxygen flow or respiration per cell [amol·s<sup>-1</sup>·x<sup>-1</sup>] = [pmol·s<sup>-1</sup>·Mx<sup>-1</sup>] is normalized for the cell count, distinguished from [[oxygen flux]] (e.g. per mg protein or wet mass). These are different forms of [[normalization of rate]].
|info=Gnaiger E et al ― MitoEAGLE Task Group (2020) Mitochondrial physiology. Bioenerg Commun 2020.1. https://doi.org/10.26124/bec:2020-0001.v1, Gnaiger E (2021) The elementary unit — canonical reviewer's comments on: Bureau International des Poids et Mesures (2019) The International System of Units (SI) 9th ed. https://doi.org/10.26124/mitofit:200004.v2, [[Gnaiger 2020 BEC MitoPathways]], [[Gnaiger_1993_Pure Appl Chem]], [[Renner_2003_Biochim_Biophys_Acta]]
}}
 
{{Template:Keywords: Normalization}}
 
{{MitoPedia concepts
|mitopedia concept=MiP concept, Ergodynamics
}}
{{MitoPedia methods
|mitopedia method=Respirometry
}}
{{MitoPedia O2k and high-resolution respirometry
|mitopedia O2k and high-resolution respirometry=DatLab
}}
}}
{{Labeling}}

Latest revision as of 10:25, 8 February 2024


                  


O2k-Open Support

Oxygen flow



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Description

Respiratory oxygen flow is the oxygen consumption per total system, which is an extensive quantity. Flow is advancement of a transformation in a system per time [mol·s-1], when 'system' is defined as the experimental system (e.g. an open or closed chamber). Flow is distinguished from the size-specific quantity flux obtained by normalization of flow per volume of the experimental system [mol·s-1·m-3]. An experimental object, e.g. a living cell, may be considered as the 'experimental system'. Then oxygen flow per cell has the unit [mol·s-1·x-1], where [x] is the elementary unit for a count. Oxygen flow or respiration per cell [amol·s-1·x-1] = [pmol·s-1·Mx-1] is normalized for the cell count, distinguished from oxygen flux (e.g. per mg protein or wet mass). These are different forms of normalization of rate.

Abbreviation: IO2 [mol·s-1] or [mol·s-1·x-1]

Reference: Gnaiger E et al ― MitoEAGLE Task Group (2020) Mitochondrial physiology. Bioenerg Commun 2020.1. https://doi.org/10.26124/bec:2020-0001.v1, Gnaiger E (2021) The elementary unit — canonical reviewer's comments on: Bureau International des Poids et Mesures (2019) The International System of Units (SI) 9th ed. https://doi.org/10.26124/mitofit:200004.v2, Gnaiger 2020 BEC MitoPathways, Gnaiger_1993_Pure Appl Chem, Renner_2003_Biochim_Biophys_Acta