Amplex UltraRed: Difference between revisions
From Bioblast
No edit summary |
No edit summary |
||
Line 146: | Line 146: | ||
Source: Life Technologies (former Invitrogen) A36006 | Source: Life Technologies (former Invitrogen) A36006 | ||
::* Krumschnabel G, Fontana-Ayoub M, Sumbalova Z, Heidler J, Gauper K, Fasching M, Gnaiger E (2015) Simultaneous high-resolution measurement of mitochondrial respiration and hydrogen peroxide production. Methods Mol Biol 1264:245-61. [[Krumschnabel 2015 Methods Mol Biol |»Bioblast pdf«]] | ::* Krumschnabel G, Fontana-Ayoub M, Sumbalova Z, Heidler J, Gauper K, Fasching M, Gnaiger E (2015) Simultaneous high-resolution measurement of mitochondrial respiration and hydrogen peroxide production. Methods Mol Biol 1264:245-61. [[Krumschnabel 2015 Methods Mol Biol |»Bioblast pdf«]] | ||
::* Makrecka-Kuka M, Krumschnabel G, Gnaiger E (2015) High-resolution respirometry for simultaneous measurement of oxygen and hydrogen peroxide fluxes in permeabilized cells, tissue homogenate and isolated mitochondria. Biomolecules 5:1319-38. [[Makrecka-Kuka 2015 Biomolecules|»Bioblast pdf«]] | |||
'''O2k-Demo experiments''' | '''O2k-Demo experiments''' |
Revision as of 16:33, 9 December 2015
Description
Amplex red (AmR) is used as an extrinsic fluorophore for measurement of hydrogen peroxide production (ROS) by cells or mitochondrial preparations. The reaction of H2O2 and AmR is catalyzed by horseradish peroxidase to produce the red fluorescent compound resorufin (excitation wavelength 563 nm, emission 587 nm). The change of emitted fluorescence intensity is directly proportional to the concentration of H2O2 added, whereby the H2O2 is consumed.
Abbreviation: AmR
Reference: Mohanty 1997 J Immunol Methods, Zhou 1997 Anal Biochem, Mishin 2010 Free Radical Biol Med, Towne 2004 Anal Biochem, Krumschnabel 2015 Methods Mol Biol
MitoPedia methods:
Fluorometry
Signal and output
- Signal: The O2k-Fluo LED2-Module is operated through the amperometric (Amp)-Channel of the O2k, with electric current (ampere [Amp]) as the primary signal.
- Output: The focus of the output with Amplex red is on Type B: Flow, flux, rate.
Different brands
Trade Mark | Manufacturer/ Distributor | price [€/mg]
|
product id
|
Amplex Red | 37.4
|
A12222
| |
Amplex Ultra Red | 48.8
|
A36006
| |
Ampliflu Red | 18.5
|
90101
| |
Quanta Red | 15150
|
Citation | Amplex | HRP | pH | Limit of detection | |||
stock | final | unit definition | stock | final | |||
mM
|
µM
|
U/ml
|
U/ml
|
||||
BIOTEK | 10
|
50
|
pyrogallol
|
10
|
0.1
|
7.4
|
4 nM (absorption 300 nm)
|
Life Technologies | 10
|
50
|
10
|
0.1
|
6 to 7.5
|
<80 nM
| |
Towne 2004 | 160
|
0.41
|
7.5 to 8.5
|
100 nM
| |||
Zhou1997 | 3 ?
|
0.3 to 1
|
50 nM (10 nM optimal)
| ||||
Mohanty 1997 | 100
|
10 to 100
|
1
|
18 nM
| |||
Komary 2010 | 1
|
2.5
|
Applications with O2k-Fluorometry
Source: Life Technologies (former Invitrogen) A36006
- Krumschnabel G, Fontana-Ayoub M, Sumbalova Z, Heidler J, Gauper K, Fasching M, Gnaiger E (2015) Simultaneous high-resolution measurement of mitochondrial respiration and hydrogen peroxide production. Methods Mol Biol 1264:245-61. »Bioblast pdf«
- Makrecka-Kuka M, Krumschnabel G, Gnaiger E (2015) High-resolution respirometry for simultaneous measurement of oxygen and hydrogen peroxide fluxes in permeabilized cells, tissue homogenate and isolated mitochondria. Biomolecules 5:1319-38. »Bioblast pdf«
O2k-Demo experiments