SUIT-004

From Bioblast


high-resolution terminology - matching measurements at high-resolution


SUIT-004

Description

1PM;2D;3U;4S;5Rot.png

Abbreviation: RP1-short

SUIT protocol pattern: 1PM;2D;3U;4S;5Rot-

The SUIT-004 protocols are designed to provide a quick assessment of the linear coupling control (L- P- E) with NADH-linked substrates (PM) and the contribution of the S-pathway to the ET state (N, NS, S). SUIT-004 can be extended with the CIV assay module.

Communicated by Doerrier C, Gnaiger E (last update 2024-08-09)

Specific SUIT protocols

1PM;2D;2c;3U;4S;5Rot;6Ama;7AsTm;8Azd.png SUIT 004.png

MitoPedia: SUIT

Steps and respiratory states

1PM;2D;2c;3U;4S;5Rot;6Ama;7AsTm;8Azd.png

Step State Pathway Q-junction Comment - Events (E) and Marks (M)
1PM PML(n) N CI 1PM
2D PMP N CI 1PM;2D
2c PMcP N CI 1PM;2D;2c
3U PME N CI 1PM;2D;3U
4S PMSE NS CI&II 1PM;2D;3U;4S
  • Respiratory stimulation by simultaneous action of type N substrates & succinate, with convergent electron flow in the NS-pathway for reconstitution of TCA cycle function.
  • Noncoupled electron transfer state, ET state, with ET capacity E.
5Rot SE S CII 1PM;2D;3U;4S;5Rot
6Ama ROX 1PM;2D;3U;4S;5Rot;6Ama
  • Rox is the residual oxygen consumption in the ROX state, due to oxidative side reactions, estimated after addition of antimycin A (inhibitor of CIII). Rox is subtracted from oxygen flux as a baseline for all respiratory states, to obtain mitochondrial respiration (mt).
Step Respiratory state Pathway control ET-Complex Comment
## AsTm AsTmE CIV CIV
## Azd CHB


Questions.jpg


Click to expand or collaps
Bioblast links: SUIT protocols - >>>>>>> - Click on [Expand] or [Collapse] - >>>>>>>

Strengths and limitations

  • A succinate concentration of >10 mM may be required for saturating SE capacity.
+ Linear coupling control (L- P- E) with N substrates (PM).
+ Pathway control in ET state (N, NS, S).
+ This protocol can be extended with the Complex IV module.
+ Short experimental time.
- The substrate combination for maximum ET capacity (FNSGp) is not assayed in SUIT-004 in favour of reducing the experimental time.

Compare SUIT protocols

  • SUIT-001 (RP1): SUIT-004 is a shortened version of RP1; without G, Oct and Gp.

References

MitoPedia concepts: MiP concept, SUIT protocol 


MitoPedia methods: Respirometry 