== O2k-Open Support: technical help ==


[[File:MitoPedia.jpg|left|100px|MitoPedia]]
:::: '''O2k-Open Support''' includes [[Open Access]] pages that help to optimally use the O2k, to apply instrumental O2k-Quality Control, and perform user-service when needed. Support regarding data analysis and/or interpretation is provided on an individual basis, for which consultancy fees apply. Β  Β  Β  Β  Β 
Β 
::::::# '''MitoPedia: O2k''': find information quickly with MitoPedia, the Bioblast glossary with lists of key words and links to relevant pages.
== MitoPedia ==
::::::## '''[[MitoPedia: O2k-Open Support]]'''
Β 
::::::## [[MitoPedia: O2k hardware]]
Find information quickly with '''[[MitoPedia]]''', the Bioblast glossary with lists of key words and links to relevant pages, specifically and the O2k and high-resolution respirometry:
::::::## [[MitoPedia: DatLab]]
<br />
::::::# [[O2k-Procedures]]: explain various applications of the O2k with O2k-Demo experiments
<br />
::::::# [[O2k-Manual]]: topic specific manuals with step by step help
::Β» [[MitoPedia: O2k hardware]]
::::::# [[O2k-Videosupport]]: provides valuable assistance, complementary to the O2k-Manual
::Β» [[MitoPedia: DatLab]]
::::::# [[O2k-Products]]: is not merely a 'List of products', it presents a guided tour to the O2k and integrates the concept of [[Gentle Science| Gentle Science and Open Science]].
::Β» [[MitoPedia: SUIT protocols]]
::Β» [[MitoPedia: SUIT states]]
Β 
Β 
[[Image:O2k-Protocols.jpg|left|100px|link=http://wiki.oroboros.at/index.php/O2k-Protocols |O2k-Protocols]]
Β 
== O2k-Protocols ==
'''O2k-Protocols''' explain various applications of the O2k with O2k-Demo experiments, SOPs beyond the O2k-Manual, and more: '''Β» [[O2k-Protocols]]'''
Β 
:* Implement your quality control management with the [[MitoFit Quality Control System |MitoFit QCS]], to diagnose technical problems fast and cost-effectively with instrumental quality control tests.
::Β» O2k-standard operating procedures: Β» [[O2k-SOP]]
::Β» Polarographic oxygen sensors: calibration, accuracy and quality control SOP. Β» [[MiPNet06.03 POS-Calibration-SOP]]
::Β» Instrumental background correction, accuracy of oxygen flux and SOP. Β» [[MiPNet14.06 InstrumentalO2Background]]
Β 
Β 
[[Image:O2k-Manual.jpg|left|100px|link=http://wiki.oroboros.at/index.php/O2k-Manual |O2k-Manual]]
Β 
== O2k-Manual ==
Β 
The '''O2k-Manual''' helps beginners and experienced users. Find technical information and instructions in the main sections (Bioblast pdf files), complemented by details on specific webpages: '''Β» [[O2k-Manual]]'''
Β 
=== O2k-Videosupport ===
::Β» '''[[O2k-Videosupport]]'''
Β 
Β 
[[Image:O2k-Catalogue.jpg|left|100px|link=http://wiki.oroboros.at/index.php/OROBOROS_O2k-Catalogue|O2k-Catalogue]]
Β 
== O2k-Catalogue ==
Β 
The O2k-Catalogue is a guided tour through the O2k technology, with links to specific sections of the O2k-Manual, and more: '''Β» [[OROBOROS O2k-Catalogue]]'''
Β 
Β 
[[Image:O2k-support system.jpg|right|300px|OROBOROS support system]]
== Open innovation ==
Β 
::* '''Open innovation''' extends the support given to individual users to the scientific community and provides feedback to improve the scope and documentation of the O2k-technology - in the spirit of [[Gentle Science]].
::* We welcome and value feedback. Technical and scientific questions, which are not yet addressed in the available guidelines or are difficult to trace on our website, will be relevant to both specialists and beginners in the field.
::* Expansion of [[O2k-SOP]]s is particularly relevant for novel approaches in O2k-Fluorometry and O2k-Spectrophotometry, which are combined in the development of the [[NextGen-O2k]].
Β 
: <big>'''Extended O2k-technical support'''</big>
::* O2k-technical support is defined as '''extended''' if support is requested on O2k-technology and high-resolution respirometry without prior consulting the [[O2k-Manual]] and relevant chapters on [[O2k-Protocols]].
::* Regular consultancy fees apply for '''extended O2k-technical support''' requested without open innovation agreement.
::* Acceptance of '''extended O2k-technical support''' with an '''open innovation agreement''' implies that OROBOROS INSTRUMENTS may publish the relevant correspondence or a summary of the correspondence addressing technical and scientific questions on our website, including names, institutional and Email addresses.
Β 
Β 
Β 
[[Image:O2k-Support.jpg|left|100px|OROBOROS support]]
== O2k-technical support ==
Β 
* '''Application problems'''
:: An '''application problem''' can be diagnosed on the basis of O2k-calibration and instrumental background tests performed in incubation medium in the absence of an experimental sample. Analysis of DatLab files recorded in such instrumental tests is essential. If proper instrumental performance is obtained, software and hardware problems are excluded.
Β 
* '''Software problems'''
:: Communication between the O2k and DatLab running on a computer.
Β 
* '''Hardware problems'''
:: Diagnostic tests are available to identify a defective instrumental component. The two-chamber design of the O2k provides the option to exchange parts between the two chambers and thus locate a defective component.
Β 
# Hardware problems that can be solved by user-service: The [[OroboPOS]] (polarographic oxygen sensor) requires user-service at intervals (these may be more than 1 year), and SOPs are available to determine if a sensor service is required at an earlier or later date.
# Hardware problems that require replacement of a defective part.
# Some electronic or mechanical defects may be solved only by service of the O2k in the workshop of OROBOROS INSTRUMENTS, e.g. a defective Peltier unit.
Β 
* The '''O2 sensor and O2k chamber tests''' are important to eliminate instrumental artefacts, which then can be distinguished from problems arising from solution protocols (perparation of chemicals), sample preparation protocols, and SUIT protocols. Biological experiments are not suitable for trouble shooting. Clean O2k-Chambers and contamination-free incubation medium are required. The sequence of quality control tests is strictly
# O2 sensor test
# O2 chamber test
Β 
* The '''two chamber design''' of the O'''2k''' has many benefits, including advantages in '''trouble shooting'''. By switching components between sides A and B of the O2k, it is usually possible to locate an instrumental problem, finding selectively a specific defective component of the O2k. This component can then be serviced by the end user or replaced by OROBOROS without the need to ship the entire O2k.
Β 
[[File:O2 sensor test.jpg|right|400px|link=MiPNet06.03 POS-Calibration-SOP |O2 sensor test]]
=== O2 sensor test ===
Many technical problems of [[O2k-Core]] applications can be analyzed and solved with the '''O2 sensor test''', described in detail as an O2k-SOP:
::Β» O2 sensor test: [[MiPNet06.03 POS-Calibration-SOP]]
Β 
* The O2 sensor test is not only required to evaluate the function of the OroboPOS. To obtain a high-resolution oxygen signal, many other components of the O2k have to function according to specifications. Examples for parts that may be not properly handled, assembled or be defective, and which may explain problems observed at the level of an O2 sensor test:
# [[DatLab]] not operated properly.
## Check [[Scaling]].
## Check settings for [[Gain]] (use Gain 1) and [[Polarization voltage]] (use 800 mV).
## If the signal remains off scale (9.99 V) or very low (< 1 V) at air saturation and 37 Β°C, many components may be responsible, and an electronic defect of the [[Main unit]] must be carefully excluded.
## Empty the chamber with the O2k running and connected to DatLab. Switch the stirrer off. Remove the [[OroboPOS connector]] with the attached [[OroboPOS]]. Leave the sensor attached to the POS connector, and the POS connector plugged into the O2k-Main Unit. Set an event and record the signal for a short time (some min). The raw signal should normalize to a value >1 V and < 3 V (Gain 1). If so, the O2k-Chamber assembly was problematic (application problem). If not, the OroboPOS must be serviced, to exclude a hardware problem of the sensor.
## Remove the sensor head from the sensor connector, which remains plugged into the O2k-Main Unit. The raw signal should be stable at 0 V. If so, the OroboPOS must be serviced, to exclude a hardware problem of the sensor. If not, a defect of the [[OroboPOS-Connector]] is indicated.
## Remove the plug of the [[OroboPOS connector]] from the [[O2k-Main unit]]. For O2k-Series D and higher: typically 0 V; for O2k-Series A to C: the typical signal is off-scale (+/- 9.99 V). A gain-independent stable signal of 0.4 V is a strong indication of an electronic problem in the O2k-Main unit.
# [[USB-Cable 2.0\Type A-B]] not properly connected to the O2k and PC or Laptop.
# [[O2k-Chamber]] not properly positioned, such that O2 sensors are not connected to the medium.
# [[OroboPOS-Membranes]] defective or not properly applied.
# [[OroboPOS-Connector]] contaminated gold contacts; plugs not properly connected to the sockets of the [[O2k-Main Unit]].
# [[Pen-Contact Oil]] OroboPOS contacts not cleaned.
# [[OroboPOS-Electrolyte Powder]] contaminated, inappropriate water used for dissolving the powder.
# [[O2-Zero Powder]] not properly handled; confused with [[OroboPOS-Polishing Powder]].
# [[OroboPOS]] not properly serviced; not properly mounted to the [[OroboPOS-Connector]]; or defective POS head.
# Room temperature not sufficiently stable.
# [[Stirrer-Bar\white PVDF\15x6 mm]] not added to the chamber, or stuck and not rotating.
# [[O2k-Barometric Pressure Transducer]] not properly calibrated.
# [[O2k-Peltier Temperature Control]] defective electronics.
# [[O2k-Electromagnetic Stirrer Twin-Control]] defective electronics.
# [[O2k-Main Basic]] not properly connected; other defective hardware.
Β 
: If a problem occurs: Stop and start '''trouble shooting''' at the level of components which secure the quality of the O2 sensor signal and O2 concentration.
# Switch the [[OroboPOS]] between O2k-Chamber A and B. This is a good opportunity to clean the gold contacts and apply [[Contact oil]] to the gold pin and thread connecting the [[POS connector]]s and sensors. It is not necessary to remove the mounted membrane and the seal tip from the OroboPOS. When disconnecting a sensor from the POS connector observe the guidelines to prevent damage by ESD ([[MiPNet14.01 ESD Damage]]).
# Repeat the O2 sensor test. If the problem moved together with the sensor from one side to the other, the problem is located in the [[OroboPOS]] sensors.
# If the problem remained on the same side, switch the [[POS connector]] together with the attached sensors between right and left side (A and B).
# Repeat the O2 sensor test. If the problem moved together with the POS connector, it is located at the [[POS connector]].
# An electronic defect in the [[Main unit]] can be detected at the level of the O2 sensor test. This should be confirmed by removing components as described below.
Β 
Β 
=== O2k-chamber test ===
If all quality control criteria of the O2 sensor test are met, the operator can be assured that the quality of the sensor signal is acceptable. Next, the quality of the O2k-Chamber assembly has to be tested.
::Β» O2k chamber test: [[MiPNet14.06 InstrumentalO2Background]]
Β 
* Examples for parts that may be not properly handled, assembled or be defective, and which may explain problems observed at the level of an O2k-chamber test, but not with the O2 sensor test:
# [[O2k-Chamber]] not properly assembled or broken.
## [[OroboPOS-Holder]] not properly positioned.
## [[O2k-Chamber Holder]] not properly positioned; V- and O-rings not properly mounted ([[V-ring\30-35-4.5 mm]], [[O-ring\Viton\18x2 mm]]).
# [[Volume-Calibration Ring]] not properly positioned by chamber volume calibration.
# [[O-ring\Viton\12x1 mm]] injured and must be replaced on the stopper.
# [[Stopper\black PEEK\conical Shaft\central Port]] broken conical edge or O-ring not properly applied.
# [[OroboPOS-Seal Tip]] leaky.
# Experimental medium consumes oxygen due to microbial contamination.
Β 
: If a problem occurs: Stop and start '''trouble shooting''' at the level of components which secure the quality of the O2 chamber and O2 flux.
# Check the stirring bars for any contamination.
# Check the stoppers for the quality of the O-rings and the conical edges.
# If no indications of a defect are observed, disassemble the O2k-Chamber.
# Check the [[glass chamber]] for contamination or for broken edges.
# Clean the copper block of the O2k and reassemble the O2k-Chamber.
# Reassemble and clean the chambers: [[MiPNet19.03 O2k-cleaning and ISS]].
# Perform an O2 sensor test and - if successful - an O2k-chamber test, using fresh incubation medium.
# If the problem with the instrumental O2 background remains in one chamber, switch stoppers between chambers A and B.
# Perform an O2k-chamber test (the sensor test is not necessary).
# If the problem with the instrumental O2 background remains in the same chamber, switch glass chambers between the left and right side of the O2k.
# Perform an O2 sensor test and - if successful - an O2k-chamber test.
# If the problem with the instrumental O2 background remains in the same chamber, switch sensors between the left and right chamber.




{{#ask: mainlabel=Title|[[Technical service::+]]
::: '''<big>O2k-Quality Control</big>'''
| ?Technical service
[[File:MitoFit-QCS.jpg|right|400px|MitoFit-QCS]]
| limit = 200
:::: Many commonly raised issues of O2k-technical support are solved in the newest version of [[DatLab]].Β  Β  Β 
}}


:::: The '''O<sub>2</sub> sensor and O2k chamber tests''' are important to eliminate instrumental artefacts (instrumental quality control, QC), distinguished from assay QC addressing problems of [[solution protocols]] (preparation of chemicals), [[sample]] preparation protocols (such as [[mitochondrial preparations]]), and [[SUIT protocols]].
::::# Clean O2k-chambers and contamination-free incubation medium are required.
::::# Analysis of DatLab files recorded in instrumental QC tests is essential.Β  Β  Β  Β 
::::# Biological experiments are not suitable for trouble shooting.


== Cost estimation for repair ==


:: A cost estimation for repair of the O2k-Main unit due to anΒ  electronic/mechanical defect can only be given after inspection of theΒ  O2k by our workshop in Austria. Once a cost estimation has been provided, the repair is carried out after the customer’s consent. For shipment of the O2k, follow the [[Shipping an oxygraph |detailed instructions]].
=== 1. O<sub>2</sub> sensor test ===
:::: The [[OroboPOS|OroboPOS]] (polarographic oxygen sensor) requires regular service at intervals - these may be more than 1 year. SOPs are available to determine if a sensor service is required at an earlier or later date:
::::Β» O<sub>2</sub> sensor test: [[Oxygen_sensor_test |'''QC1: Oxygen sensor test''']] Β  Β  Β 


=== 2. O2k-chamber test ===
:::: If all quality control criteria of the O<sub>2</sub> sensor test are met, the operator can be assured that the quality of the sensor signal is acceptable. Next, the quality of the O2k-chamber assembly has to be tested, described in detail:
::::Β» O2k-chamber test: [[Oxygen flux - instrumental background |'''QC2: Instrumental O<sub>2</sub> background test''']]Β  Β  Β 


== OROBOROS MitoFit Laboratory ==


== O2k-Open Support: Installation and startup support session ==
:::::Β» '''[[MiPNet26.07 Installation and startup support session]]'''


:: The '''[[OROBOROS MitoFit Laboratory]]''' is open for innovation and cooperation.


:::: Upon special request and subject to an additional fee, Oroboros Instruments offers a start-up visit by a scientific application specialist.
:::::Β» '''[[O2k On-site Training]]'''


== The O2k-Network - are you connected? ==


:: The OROBOROS O2k-Network serves to connect and support:
[[File:O2k-Support.jpg|right|200px|O2k-Open Support]]
== O2k-Open Support agreement ==


:::: The Oroboros experts will be happy to help in using our O2k-Open Support system.
::::Β» [[Search for defective O2k components]]
:::: '''Β»''' In case you need further technical support and for your valuable feedback, please contact [https://www.oroboros.at/index.php/o2k-technical-support '''O2k-Tech Support''']Β  Β  Β 
::::Β» [[O2k-Open Support agreement]]
::::Β» [[O2k repair]]


[[Image:O2k-Network.png|left|120px|link=O2k-Network|O2k-Network]]
::* '''Β»''' The O2k-Network Labs: [[O2k-Network |Β» O2k-Network]].
::* '''Β»''' Communication on a scientific level: [[O2k versus multiwell respirometer]]
::* '''Β»''' Studies in mitochondrial physiology: [http://www.oroboros.at/?high-resolution-respirometry Β» Applications].
::* '''Β»''' [[OROROBOS Science Scholarship]]s


[[File:O2k-Network.png|left|100px|O2k-Network|link=O2k-Network]]
== The O2k-Network ==


:::: The Oroboros O2k-Network serves to connect and support: Contact for advice on applications of the O2k and high-resolution respirometry - the global network in mitochondrial physiology.Β  Β  Β 


== Contact ==
::::: '''Β»''' The O2k-Network Labs: [[O2k-Network|Β» O2k-Network]]Β  Β  Β 


:: In case you need further technical support, contact [mailto:support@oroboros.at| support@oroboros.at]
::::: '''Β»''' [[O2k-Network_discussion_forum|O2k-Network discussion forum]] Β  Β  Β 


:::: '''Β»''' [[Oroboros Laboratory|Oroboros O2k-Laboratory]] - open for innovation and cooperation.
:::: '''Β»''' [[Oroboros_Science_Scholarship|Oroboros Science Scholarships]] - are offered for focused research projects on mitochondrial physiology and high-resolution respirometryΒ  Β  Β 


[[Category:OroboPedia]]
[[Category:O2k-Open Support]]

Latest revision as of 16:43, 9 August 2022


                  


O2k-Open Support

O2k-Open Support


Template NextGen-O2k.jpg
Oroboros support system
Currently Asked Questions - Β»O2k-Open Support alertΒ« 

O2k-Open Support: technical help

O2k-Open Support includes Open Access pages that help to optimally use the O2k, to apply instrumental O2k-Quality Control, and perform user-service when needed. Support regarding data analysis and/or interpretation is provided on an individual basis, for which consultancy fees apply.
  1. MitoPedia: O2k: find information quickly with MitoPedia, the Bioblast glossary with lists of key words and links to relevant pages.
    1. MitoPedia: O2k-Open Support
    2. MitoPedia: O2k hardware
    3. MitoPedia: DatLab
  2. O2k-Procedures: explain various applications of the O2k with O2k-Demo experiments
  3. O2k-Manual: topic specific manuals with step by step help
  4. O2k-Videosupport: provides valuable assistance, complementary to the O2k-Manual
  5. O2k-Products: is not merely a 'List of products', it presents a guided tour to the O2k and integrates the concept of Gentle Science and Open Science.


O2k-Quality Control
MitoFit-QCS
Many commonly raised issues of O2k-technical support are solved in the newest version of DatLab.
The O2 sensor and O2k chamber tests are important to eliminate instrumental artefacts (instrumental quality control, QC), distinguished from assay QC addressing problems of solution protocols (preparation of chemicals), sample preparation protocols (such as mitochondrial preparations), and SUIT protocols.
  1. Clean O2k-chambers and contamination-free incubation medium are required.
  2. Analysis of DatLab files recorded in instrumental QC tests is essential.
  3. Biological experiments are not suitable for trouble shooting.


1. O2 sensor test

The OroboPOS (polarographic oxygen sensor) requires regular service at intervals - these may be more than 1 year. SOPs are available to determine if a sensor service is required at an earlier or later date:
Β» O2 sensor test: QC1: Oxygen sensor test

2. O2k-chamber test

If all quality control criteria of the O2 sensor test are met, the operator can be assured that the quality of the sensor signal is acceptable. Next, the quality of the O2k-chamber assembly has to be tested, described in detail:
Β» O2k-chamber test: QC2: Instrumental O2 background test


O2k-Open Support: Installation and startup support session

Β» MiPNet26.07 Installation and startup support session


Upon special request and subject to an additional fee, Oroboros Instruments offers a start-up visit by a scientific application specialist.
Β» O2k On-site Training


O2k-Open Support

O2k-Open Support agreement

The Oroboros experts will be happy to help in using our O2k-Open Support system.
Β» Search for defective O2k components
Β» In case you need further technical support and for your valuable feedback, please contact O2k-Tech Support
Β» O2k-Open Support agreement
Β» O2k repair


O2k-Network

The O2k-Network

The Oroboros O2k-Network serves to connect and support: Contact for advice on applications of the O2k and high-resolution respirometry - the global network in mitochondrial physiology.
Β» The O2k-Network Labs: Β» O2k-Network
Β» O2k-Network discussion forum
Β» Oroboros O2k-Laboratory - open for innovation and cooperation.
Β» Oroboros Science Scholarships - are offered for focused research projects on mitochondrial physiology and high-resolution respirometry