Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "MiPNet06.03 POS-calibration-SOP"

From Bioblast
 
(106 intermediate revisions by 16 users not shown)
Line 1: Line 1:
{{OROBOROS navigation line page name}}
{{Template:OROBOROS support page name}}
{{Publication
{{Publication
|title=[[Image:O2k-Protocols.jpg|right|80px|link=O2k-Protocols|O2k-Protocols]] Polarographic oxygen sensors: calibration, accuracy and quality control SOP. [[Media:MiPNet06.03 POS-Calibration-SOP.pdf |»Bioblast pdf«]]
|title=[[Image:O2k-Manual.jpg|right|70px|link=O2k-Manual|O2k-Manual]]O2k Quality Control 1: Polarographic oxygen sensors and accuracy of calibration.
|info=[http://www.bioblast.at/index.php/File:MiPNet06.03_POS-Calibration-SOP.pdf Versions]
|info=[[File:PDF.jpg|100px|link=https://wiki.oroboros.at/images/7/77/MiPNet06.03_POS-Calibration-SOP.pdf |Bioblast pdf]] » [http://www.bioblast.at/index.php/File:MiPNet06.03_POS-Calibration-SOP.pdf Versions]
|authors=OROBOROS
|authors=Oroboros
|year=2015-05-10
|year=2023-02-06
|journal=Mitochondr Physiol Network
|journal=Mitochondr Physiol Network
|abstract=Gnaiger E (2015) Polarographic oxygen sensors: calibration, accuracy and quality control SOP. Mitochondr Physiol Network 06.03(13):1-8.
|abstract=Gnaiger E (2023) O2k Quality Control 1: Polarographic oxygen sensors and accuracy of calibration. Mitochondr Physiol Network 06.03(21):1-10.
 
{{MiPNet pdf page linking to MitoPedia}}
[[High-resolution respirometry]] critically depends on accurate calibration of the [[polarographic oxygen sensor]]s which respond to [[Oxygen_pressure| partial oxygen pressure]].  Expressing the oxygen signal in terms of partial pressure has several advantages, but respiration is expressed in molar units related to biochemical stoichiometries. Conversion of oxygen partial pressure to [[oxygen concentration]] requires accurate information on oxygen solubilities in experimental incubation media. Absolute errors up to 10% are commonly found in the established literature. The [[oxygen solubility]] of mitochondrial respiration medium [[MiR06]] relative to pure water (oxygen solubility factor, ''F''<sub>M</sub>) is 0.92, accurately determined (for MiR05, i.e. MiR06 without catalase; [[MiPNet14.13]]) at 37 °C and 30 °C.  At air saturation at standard barometric pressure (100 kPa) and 37 °C, the partial oxygen pressure is 19.63 kPa, and the oxygen concentration is 190.7 µM in MiR06.  A clean and ready-to-use O2k-Chamber is a requirement for saving time and quickly calibrating the polarographic oxygen sensors.
|mipnetlab=AT_Innsbruck_Oroboros
:» Product: [[OroboPOS]], [[Oxygraph-2k]], [[O2k-Catalogue_OROBOROS |O2k-Catalogue]]
|mipnetlab=AT_Innsbruck_OROBOROS
}}
}}
{{Labeling
|area=Instruments;methods
|instruments=Oxygraph-2k, Protocol
|additional=O2k-SOP
}}
== This article is cited in ==


* Rodriguez JV, Pizarro MD, Scandizzi AL, Guibert EE, Almada LL, Mamprin ME (2008) Construction and performance of a minibioreactor suitable as experimental bioartificial liver. Artif Organs 32:323-8. [http://www.ncbi.nlm.nih.gov/pubmed/18370948 »PubMed 18370948«]
== Keywords ==
{{Template:Keywords: Oxygen signal}}


::::» [[Oxygen calibration - DatLab]]


== Further details ==
* O2k-Manual: »[[MiPNet19.18D_O2k-Calibration]]«
* Next step of instrumental quality control: »[[MiPNet14.06 InstrumentalBackground]]«
* Forstner H, Gnaiger E (1983) Calculation of equilibrium oxygen concentration. In: Polarographic Oxygen Sensors. Aquatic and Physiological Applications. Gnaiger E, Forstner H (eds), Springer, Berlin, Heidelberg, New York:321-33. [[Forstner_1983_POS |»Bioblast Access«]]


 
{{Labeling
[[Image:MiPNet06.03_Table.jpg|centre|frame|<sup>c</sup>O2* is the oxygen concentration at thermodynamic equilibrium with air under standard conditions of 100 kPa barometric pressure. ''F''<sub>M</sub> has been determined for MiR05 at 30 °C and 37 °C, and is assumed to be identical at 25 °C.]]
|area=Respiration, Instruments;methods
 
|instruments=Oxygraph-2k, O2k-Manual
== O2 sensor test ==
|additional=DatLab, O2k-SOP, DL7, Oxygen solubility, DL8,  
{{Technical support integrated}}
MitoFit 2021 Dark respiration, MitoFit 2021 Photosynthesis, DL8, PLoSONE2022ace-sce, MitoFit2022QC, POS
{{#set:Technical service=O2 sensor|Technical service=POS connector|Technical service=O2 signal|Technical service=Main unit|Technical service=Performance test}}
}}
'''The O<sub>2</sub> sensor test is described as an [[OroboPOS]]''' standard operating procedure: »Bioblast pdf (on top of page).
 
'''When should an O2-sensor test be performed?'''
* After switching on the O2k, every day: air calibration and stirrer test.
* Zero oxygen calibration: from time to time over weeks; bracketing zero oxygen calibrations when working at low oxygen.
* After application of a new membrane and O2-sensor service: In some cases, the signal of the OroboPOS improves (higher signal stability, less noise, shorter response time), when the O2k remains switched on over night (O2k-Chambers filled with 70% EtOH).
* During troubleshooting, particularly when [[switching components]] between the two chambers, a quick sensor test is performed after each step (stirrer test, sensor signal).
 
'''Main criteria for sensor performance'''
=== Oxygen signal at air saturation ===
The oxygen signal at air saturation is an important performance criterion. It is assessed in the sensor test during the preparation for each experiment.
 
The uncalibrated oxygen sign (raw signal) depends on the oxygen concentration and the [[gain]]. The numerical value of the calibrated oxygen signal depends on the raw signal and the [[POS calibration]]. The optical appearance of the calibrated oxygen signal also depends on the used [[scaling]].
 
If there are any doubts about the oxygen signal, please check first [[Raw signal at air saturation]]. If the raw signal is in the expected range for the chosen gain and stable the problem is probable a software setting.
In this case check that
* the gain is set correctly (as described in [[Raw signal at air saturation]]
* a proper [[POS calibration]] has been performed for this chamber with this sensor at this [[gain]]. If necessary repeat [[air calibration]] and [[zero calibration]]
 
If the oxygen signal at air saturation, R1 [ [[raw signal]] ; V], is suddenly very different compared to previous calibrations see also [[Suddenly changing oxygen signal]].
 
For a discussion about the flux at air saturation see also [[Flux at closed chamber near to air saturation]] and [[Flux at open chamber]].
 
* [[Zero current |Raw signal at zero oxygen]]
* [[Signal stability]]: drift and noise
 
 
== OroboPOS technical support ==
{{Technical support integrated}}
If a [[Stirrer test]] shows a slow response of the sensor, see more details in [[Slow responding sensor]].
 
As the heart of the OROBOROS O2k the OroboPOS is the prime suspect when technical problems are encountered. However, frequently problems can be traced either to inadequate software settings or to other hardware components, such as the [[POS connector (technical service) |POS connector]]). Therefore, it is important to [[Locating a problem |locate the problem]], primarily with a [[Sensor test]].
 
=== OroboPOS technical support pages ===
{{#ask: mainlabel=Title|[[Technical service::O2 sensor||O2 signal ]]| ?Technical service}}
 
 
=== Problems that require a full Sensor test to locate the problem ===
* Unstable O2 signal.
* Unstable O2 flux.
* High O2 signal at air saturation.
* High O2 signal at zero oxygen.
 
=== Special cases: no further localization necessary ===
* [[Slow responding sensor]], if observed in a [[Stirrer test]]. Note: The stirrer test is part of the sensor test. It is therefore recommended to do a full [[Sensor test]] whenever problems are encountered.
 
=== General trouble shooting procedure for the OroboPOS ===
 
Confirm the problem by doing a [[Sensor test]] in the absence of biological material
 
==== Case 1.A: No problem was visible in the [[Sensor test]] ====
Check the following DatLab settings:
* [[Gain (O2 channel)]]
* [[POS calibration]]
* Scaling
 
If the problem is not visible when observing the [[Raw signal]] but is visible when observing the calibrated oxygen signal, then there is probably a problem with the [[POS calibration]].
 
==== Case 1.B: The problem was visible in the [[Sensor test]] ====
* Proceed to [[ Locating a problem|Locate the Problem]] and continue as shown below.
 
==== Case 2.A: The problem could be localized on the [[OroboPOS]] ====
'''Solutions''': Follow the instructions for oxygen sensor service [[MiPNet08.04]]. Apply [[contact oil]] to the gold pin and thread connecting the [[OroboPOS]] and [[sensor connector]]. After a sensor service or after applying a new membrane, repeating the [[sensor test]]. In some cases the O2k may be left running over night to achieve a stable signal, as seen by a repeated [[sensor test]].
If all service precautions fail to solve the problem, the [[OroboPOS]] has to be sent to OROBOROS INSTRUMENTS for further service (a recovery cannot be guaranteed), or a new sensor is applied.
 
==== Case 2.B: The problem was not located on the [[POS]] ====
'''Solutions''': If the problem was localized on an other component follow the instructions for this component. If you could not locate the problem, contact OROBOROS INSTRUMENTS.
 
 
== Response time of the POS ==
{{Technical support integrated}}
{{#set:Technical service=O2 sensor|Technical service=O2 signal}}
The response time of the [[Polarographic oxygen sensor]] is assessed in the [[Stirrer test]] that is part of a full [[Sensor test]]. A quantitative interpretation of the stirrer test will yield the time constant of the sensor. Such a quantitative calculation is indeed necessary for kinetic work. However, to access the suitability of a sensor for standard steady state work each user will soon gain the necessary experience to judge the outcome of a stirrer test.
:>> Example: Download [http://www.oroboros.at/index.php?id=sensor_test here].
 
For how to improve the response time of a POS, see
[[Slow responding sensor]].
 
The response time of the sensor is just one of several factors determining the [[time resolution]].
 
== Stirrer test ==
{{Technical support integrated}}
{{#set:Technical service=O2 sensor|Technical service=O2 signal|Technical service=Performance test}}
The [[response time of the POS]] can be most conveniently checked by a stirrer test, see
[[MiPNet06.03_POS-Calibration-SOP]], section 3.
The stirrer test is part of the [[Sensor test]] and of our proposed quality checks before every experiment described in the link above.  see  If the results of a stirrer test are not satisfactory the procedures suggested for a [[Slow responding sensor]] should be followed. DatLab supports an automatically controlled stirrer test [F9].
 
Background information:
After switching off the stirrer the electrolyte reservoir will quickly be depleted of oxygen by the oxygen consumption of the sensor. Oxygen can only reach the sensor now in a unstirred diffusion process, therefore a new and lower equilibrium O2 concentration is established in the vicinity of the sensor. After switching the stirrer on again the O2 concentration near the sensor is quickly brought back to the same level as elsewhere in the chamber, however the sensor takes some time to respond to this change. Thus the response time of the sensor can be determined. A stirrer tests does not influence the O2 concentration in the entire chamber. A stirrer test can not be at a chamber oxygen concentration of zero (or near zero) because no change in the O2 signal will result.
 
 
[[Category:Technical service]]

Latest revision as of 14:27, 27 July 2023


                  


O2k-Open Support

MiPNet06.03 POS-calibration-SOP


Publications in the MiPMap
O2k-Manual
O2k Quality Control 1: Polarographic oxygen sensors and accuracy of calibration.

» Bioblast pdf  » Versions

Oroboros (2023-02-06) Mitochondr Physiol Network

Abstract: Gnaiger E (2023) O2k Quality Control 1: Polarographic oxygen sensors and accuracy of calibration. Mitochondr Physiol Network 06.03(21):1-10.

O2k-technical support and open innovation
Open the pdf document above.
» Current O2k-series: NextGen-O2k Series XB and O2k Series J
» Current software versions DatLab 8.0: MitoPedia: DatLab


O2k-Network Lab: AT_Innsbruck_Oroboros


Keywords


Questions.jpg


Click to expand or collaps


» Oxygen calibration - DatLab


Labels: MiParea: Respiration, Instruments;methods 





HRR: Oxygraph-2k, O2k-Manual 

DatLab, O2k-SOP, DL7, Oxygen solubility, DL8, MitoFit 2021 Dark respiration, MitoFit 2021 Photosynthesis, DL8, PLoSONE2022ace-sce, MitoFit2022QC, POS